FR Recognition List Number 057 Date of Entry 12/20/2021 
FR Recognition Number 13-120
Standard
ANSI AAMI 2700-1:2019
Medical Devices and Medical Systems - Essential safety requirements for equipment comprising the patient-centric integrated clinical environment (ICE) - Part 1: General requirements and conceptual model
Scope/Abstract
This standard specifies general requirements, a model and framework for integrating equipment to create a INTEGRATED CLINICAL ENVIRONMENT (ICE), as defined in 3.6. This standard specifies the characteristics necessary for the safe integration of MEDICAL DEVICES and other equipment, via an electronic interface, from different MANUFACTURERS into a single medical system for the care of a single high acuity PATIENT. This standard establishes requirements for a medical system that is intended to have greater error resistance and improved PATIENT safety, treatment efficacy and workflow efficiency than can be achieved with independently used MEDICAL DEVICES.

This series of standards establishes requirements for design, verification, and validation processes of a modelbased integration system for an INTEGRATED CLINICAL ENVIRONMENT.

This series of standards is intended to define the requirements essential for safety and thereby facilitate regulatory acceptance. NOTE These requirements were derived to support the clinical scenarios or clinical concepts of operations described in Annex B.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Transition Period
FDA recognition of ASTM 2761-09 (2013) [Rec# 13-46] will be superseded by recognition of ANSI/AMMI 2700-1:2019 [Rec# 13-120]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 13-46] until December 18, 2022. After this transition period, declarations of conformity to [Rec# 13-46] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§870.2300 System, Network And Communication, Physiological Monitors Class 2 MSX
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005
FDA Technical Contact
 Sandy Weininger
  FDA/OC/CDRH/OSEL/DBP/
  301-796-2582
  sandy.weininger@fda.hhs.gov
Standards Development Organizations
ANSI American National Standards Institute https://www.ansi.org/
AAMI Association for the Advancement of Medical Instrumentation http://www.aami.org
FDA Specialty Task Group (STG)
Software/Informatics