• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 061 Date of Entry 12/18/2023 
FR Recognition Number 14-594
Standard
ISO  11607-1 Second edition 2019-02 [Including AMD1:2023]
Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including Amendment 1 (2023)]
Scope/Abstract
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Transition Period
FDA recognition of ISO 11607-1 Second edition 2019-02 [Rec# 14-530] will be superseded by recognition of ISO 11607-1 Second edition 2019-02 [Including AMD1:2023] [Rec# 14-594 ]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 14-530] until December 20, 2026. After this transition period, declarations of conformity to [Rec# 14-530] will not be accepted.
Relevant FDA Guidance and/or Supportive Publications*
1. ISO 11607-2 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes.

2. ISO/TS 16775 Second edition 2021-11 Packaging for terminally sterilized medical devices - Guidance on the application of ISO 11607-1 and ISO 11607-2.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Steven Turtil
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIB/
  301-796-6305
  steven.turtil@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Sterility
*These are provided as examples and others may be applicable.
-
-