| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
059
|
Date of Entry 12/19/2022
|
|
FR Recognition Number
|
6-484
|
| Standard | |
ASTM F3502-22a Standard Specification for Barrier Face Coverings |
|
Scope/Abstract1.1 This specification is intended to help ensure barrier face coverings meeting the stated requirements provide (1) a means of source control for individual wearers by reducing expelled aerosols from the wearer's nose and mouth into the air; and (2) a degree of particulate filtration that potentially reduces the amount of aerosols inhaled by the wearer. ... 1.2 This specification establishes minimum design, performance (testing), labeling, user instruction, reporting and classification, and conformity assessment requirements for barrier face coverings.
1.2.1 Design criteria include setting minimum areas of face coverage over the wearer's nose and mouth, prohibiting open vents or valves, requiring a means for retaining the barrier face covering on the wearer's head, and providing a representation of product sizing. Manufacturers are further required to perform a design analysis for assessing leakage of exhaled air from the barrier face covering. Manufacturers are permitted to conduct quantitative testing as specified in this standard to supplement the design analysis.
1.2.2 Performance and testing criteria define minimum barrier face covering filtration efficiency and airflow resistance performance properties. Sub-micron particulate filtration efficiency represents the ability to capture and reduce respirable aerosols that potentially contain viruses and bacteria. Airflow resistance represents the wearer's ease of breathing or breathability while wearing the barrier face covering. The impact of repeated cleaning or laundering on continued performance is applied for measuring performance properties for those barrier face coverings that are intended for reuse. Manufacturers are permitted to also provide test results for bacterial filtration efficiency (BFE) as supplemental information to the mandatory performance measurement of sub-micron particulate filtration efficiency. ... 1.2.3 Labelling requirements specify the minimum content for labels that appear on the barrier face covering, its immediate packaging, and if different, point-of-sale packaging.
1.2.4 User instructions are required to guide selection and sizing, proper use (positioning and adjustment), and care including cleaning or laundering if product reuse is intended; inform on product cautions and limitations; and describe product replacement and disposal procedures.
1.2.5 Conformity assessment is demonstrated following Guide F3050, Annex A3, Model A to issue a declaration of conformity indicating that each barrier face covering labelled as compliant has met all of the requirements of this standard specification including design criteria, performance criteria, test methods, labelling, and user information. Additionally, conformance to this standard requires that sub-micron particulate filtration efficiency and airflow resistance tests have been performed by a laboratory accredited for conducting these tests.
1.3 This specification addresses all barrier face coverings and only barrier face coverings that are intended for either a single use (disposable) or multiple uses (reusable). ... |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ASTM F3502-22a [Rec# 6-484] will be superseded by recognition of ASTM F3502-25 [Rec# 6-517]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 6-484] until December 19, 2027. After this transition period, declarations of conformity to [Rec# 6-484] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| N/A |
Barrier Face Covering Within The Scope Of The Face Mask And Barrier Face Covering Enforcement Policy |
|
QOZ
|
|
Relevant FDA Guidance and/or Supportive Publications*
1. 42 CFR Part 84 Approval of Respiratory Protective Devices.
2. Enforcement Policy for Face Masks, Barrier Face Coverings, Face Shields, Surgical Masks, and Respirators During the Coronavirus Disease (COVID-19) Public Health Emergency (Revised), Guidance for Industry and Food and Drug Administration Staff, Issued September 2021.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
FDA Specialty Task Group (STG)
| General Plastic Surgery/General Hospital |
|
|
|
| *These are provided as examples and others may be applicable. |