 Class 2 Device Recall BRILLIANCE CT, PRIVATE PRACTICE CV CONFIGURATION, CT SCANNERsee related information
Date Initiated by FirmJune 24, 2019
Date PostedAugust 06, 2019
Recall Status1 Terminated 3 on October 06, 2022
Recall NumberZ-2191-2019
Recall Event ID 83365
510(K)NumberK042293 
Product Classification System, x-ray, tomography, computed - Product Code JAK
Product Brilliance 64 Model # 728231, computed tomography x-ray system
Code Information System Serial Number 335, 338, 1897, 4001, 4003, 4016, 4021, 9009, 9029-9030, 9033, 9035, 9038, 9040, 9045, 9047, 9048, 9052, 9054, 9056, 9058, 9062, 9064-9068, 9070-9077, 9080, 9086, 9087,9089, 9092, 9093, 9098, 9099, 9104- 9108, 9112, 9117 9120, 9121, 9127, 9134, 9138, 9140, 9148, 9159- 9160

9171, 9187, 9201, 9208, 9234, 9243, 9501, 9509 through 9608, 9610 through 10288, 10291 through 10327, 10330 through 10339, 10341 through 10808, 10811-10812,  29002, 29005, 29007, 29009, 29012, 29014, 29016-29018, 29020,29022, 29024, 29025-29027, 29032, 29035, 29037-29039, 29044-29045, 29047- 29049 29053- 29054, 29056- 29057, 29062, 29064-29066, 29068, 29081, 29086 29092, 29097, 29102, 29107, 29108, 29111, 29115, 29118, 29124, 29135 29138, 32065, 64045, 64051, 90029- 90031, 90041, 90046, 90047, 90054, 90056, 90058, 90062, 9006390066, 90069, 90072, 90078, 90079, 90081, 90082, 90085, 90086, 90087, 90089,90091, 90094, 90097, 90098, 90107, 90111, 90119, 90123-90124, 90128, 90131, 90135, 90136 90137, 90139, 90140, 90141, 90143, 90147, 90150, 90154, 90157, 90158, 90161,90163, 90167-90170, 90175,90177,90180, 90185, 90189,90199, 90205, 90213, 90218, 90223-90225, 90227, 90804, 91003, 95003, 95004, 95008, 95009, 95013- 95014

95017 through 95026, 95028, 95030 through 95061, 95063, 95069, 95071 through 95114, 95116-95118, 95130 through 95296, 95298 through 95476, 95478-95479, 95481 through 95529, 95531 through 95554, 95556 through 95566, 95568, 95570 through 95575, 95577 through 95584, 95586, 95588-95589, 95591through 95947, 122706, 320034, 760014, 6000861, 6001993, 6002294, 6002305, 6003273, 6003368, 6003844, 6003941, 200000618-1, 10003/11205, 10081, 10182EBW13108, 1155-16, 1157-09, 1157-10, 1158-05, 1190-02, 1339-09, 1343-02, 1347-04, 1348-01, 1349-01, 1350-02, 1OO82, 4101_45, 4272-08, 9208A, 95040, 95459, 95585/11737/80302, 9944, 95611, 95561/ISP80282                         
FEI Number 1525965
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information ContactMs. Holly Wright Lee
440-483-5777
Manufacturer Reason
for Recall
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
FDA Determined
Cause 2
Other
ActionPhilips Healthcare issued a Field Safety Notification (FSN)72800720/88200522 1. Identify affected product 2. There are 3 scenarios users may experience which result in identical images that are incorrectly labeled as phase tolerance images. A. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs when Start Final Recon is selected prior to display of the ECG wave at the top of the acquisition window. B. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs if ECG leads become disconnected mid-acquisition or when the acquisition is halted prematurely due to an application crash. C. By system design, in certain instances when a patient exhibits an arrhythmia or varying heart rate during a Step and Shoot acquisition, images from the acquisition following the heart rate variation, may be labeled in the incorrect phase 4. Determine whether you have a potentially affected system, then determine your AFFECTED PRODUCTS software version. To identify the software version of your product: Click the Help button, Select About and the software version is then displayed, Software versions 4.x, 3.x, and 2.x are potentially affected. 5. Take the Action listed in the field safety notice. Please retain a copy with the equipment Instruction for use. If you need any further information or support concerning this issue, please contact your local Philips representative: North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377 and follow the prompts).
Quantity in Commerce582
DistributionWorldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Timor-Leste, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, and Yemen
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JAK
-
-