| Date Initiated by Firm | February 23, 2024 |
| Date Posted | February 11, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1123-2025 |
| Recall Event ID |
96041 |
| Product Classification |
Cabinet, table and tray, anesthesia - Product Code BRY
|
| Product | BD Pyxis MedStation Auxiliary ES REF: 324.
Intended to securely store and dispense medications to a qualified and authorized healthcare provider. |
| Code Information |
UDI: 10885403512681/ Serial Numbers: 16183869 16183877 16181308 16181309 16184719 16184720 16155214 16167998 16173711 16173716 16173724 16173725 16173726 16173727 16173728 16156524 16156529 16163412 16163413 16163416 16163417 16163418 16163420 16155945 16182183 16182295 16182298 16184712 16187332 16173521 16173523 16190551 16190552 16190553 16190554 16167592 16185663 16188550 16188551 16188560 16188561 16173437 16173439 16103949 16173522 16177055 16177056 16173680 16173681 16173714 16173717 16173718 16173719 16173721 16173722 16173723 16177910 16185638 16188602 16188603 16192625 16192634 16182181 16182185 16200812 16169401 16167449 16167996 16162741 16162739 16162742 16162743 16162744 16162745 16161320 16161322 16161324 16167522 16194353 16183841 16183843 16195336 16156521 16144089 16144090 16144095 16144096 16145983 16167581 16167583 16167585 16167587 16167589 16192307 16192309 16192622 16192627 16192631 16102579 16196727 16196729 16196626 16195229 16192630 16196649 16192240 16173615 16173616 16177914 16198683 16194363 16195232 16182179 16182180 16182182 16182292 16182296 16182297 16164498 16192308 16155092 16153374 16155093 16155086 16149567 16149568 16168891 16168893 16161075 16161077 16109954 16161082 16161083 16183881 16184147 16184177 16195332 16195334 16195338 16155966 16155967 16155968 16187307 16187312 16187314 16187318 16167351 16195841 16183868 16183879 16200806 16167457 16187702 16187716 16187718 16187720 16187722 16187725 16187732 16187736 16190308 16196664 16198551 16200827 16200828 16200831 16080977 16167445 16167446 16172514 16162529 16161071 16161073 16161076 16188224 16188225 16184150 16184162 16161349 16161350 16183849 16183850 16146702 16187738 16187739 16187742 16187743 16187744 16187745 16187746 16048532 16048533 16176335 16176336 16177014 16177020 16177021 16177026 16177041 16177042 16177043 16188975 16188978 16185612 16185613 16185616 16185619 16185620 16185622 16177830 16177833 16177835 16177836 16177837 16177842 16177844 16167650 16167340 16149509 16180596 16180597 16180598 16173623 16173624 16173625 16173627 16173628 16173629 16173630 16173631 16173632 16147475 16180532 16164505 16173753 16190555 16190556 16190558 16190560 16206967 16163221 16163222 16163223 16163224 16163225 16163226 16163227 16163228 16163229 16163230 16163231 16087652 16087653 16087654 16087655 16087656 16192325 16167463 16164506 16167464 16195850 16199855 16199856 16199858 16194317 16194318 16194323 16194324 16194325 16183810 16164531 16164532 16164537 16183815 16183817 16180657 16184788 16184790 16184791 16183990 16183992 16184000 16184021 16184022 16184024 16173475 16183821 16183838 16183842 16192234 16192235 16192245 16192268 16192271 16192292 16189049 16205773 16205771 16205705 16149176 16167514 16183999 16195271 16195272 16195273 16195274 16195275 16185654 16185656 16180609 16180610 16180613 16180614 16180616 16180617 16180619 16180620 16180628 16180630 16168877 16195836 16195843 16195844 16205766 16166735 16166736 16166742 16192867 16185605 16185606 16185608 16185627 16185629 16185631 16185633 16128470 16199219 16188617 16188618 16199962 16169210 16169212 16169214 16169215 16169216 16169217 16169218 16169219 16169220 16169223 16169224 16199847 16199849 16199850 16199851 16188638 16188639 16194355 16203920 16116410 16192653 16192654 16192655 16192656 16192657 16192658 16192659 16192660 16192661 16192662 16192663 16192664 16192665 16192666 16192667 16192668 16192669 16192670 16192671 16195902 16195908 16187706 16187707 16187652 16203877 16203879 16167442 16167443 16167447 16167448 16203769 16162740 16146656 16146657 16146660 16177856 16178058 16178062 16178068 16178072 16169221 16203876 16203925 16176951 16176956 16176960 16176963 16176965 16176967 16176969 16176972 16176975 16176977 16176979 16176981 16176983 16176987 16177002 16177009 16137660 16184184 16195903 16189657 16189659 16189662 16173558 16173560 16173585 16173589 16173592 16173593 16198489 16198447 16173626 16199133 16199137 16195339 16195340 16195341 16189045 16189051 16189080 16169213 16205760 16205772 16195383 16195384 16195385 16195387 16195388 16195389 16195390 16195392 16195393 16195394 16195396 16201029 16203918 16201036 16201038 16201041 16201046 16201048 16207871 16207873 16196647 16196648 16196650 16173520 16203881 16207875 16207876 16176330 16177017 16177018 16177019 16177022 16177024 16202515 16203167 16196689 16196690 16196701 16196702 16188645 16188646 16188656 16188658 16188659 16200943 16200962 16173715 16193873 16193874 16193878 16216759 16203915 16207878 16207880 16209695 16180599 16180603 16180605 16180608 16196718 16198681 16195914 16207882 16207883 16196651 16196652 16196653 16177863 16177874 16178053 16178054 16178056 16178055 16178057 16178059 16178060 16178065 16178066 16178070 16178073 16178076 16208227 16200862 16203929 16203931 16205243 16205244 16205250 16205252 16205261 16205266 16205267 16173587 16173594 16173595 16173597 16209639 16209640 16209641 16209642 16203851 16203852 16203863 16192034 16192035 16192036 16192037 16180632 16183948 16183949 16183950 16183951 16183953 16183955 16183956 16183958 16183959 16098982 16180561 16099044 16178110 16178111 16178112 16178113 16178114 16178115 16180573 16180574 16180575 16180589 16088405 16189711 16189713 16189723 16189726 16173524 16173526 16198448 16199135 16199138 16200873 16200964 16146487 16146488 16146489 16203835 16208261 16208262 16208269 16208270 16202505 16203166 16196705 16196706 16169317 16137631 16199908 16199910 16189130 16210726 15863156 16208263 16208265 16208267 16208272 16208279 16087271 16087274 16190349 16190350 16173559 16173583 16173584 16173588 16173590 16173591 16173596 16180602 16204875 16204881 16204885 16209000 16209002 16209003 16173763 16173766 16173767 16173770 16163235 16173774 16173781 16137722 16192241 16209004 16178055 16178067 16155122 16155123 16155124 16155125 16155126 16155127 16155128 16155129 16155130 16189041 16189043 16189044 16189046 16189047 16189059 16189061 16189078 16211337 16204981 16204983 16204986 16204989 16204990 16204991 16204995 16204996 16203772 16203775 16193841 16193842 16193852 16193853 16193854 16149510 16198365 16209697 16209057 16209058 16209059 16209060 16209064 16209065 16208380 16208381 16134288 16134290 16134292 16134301 16205000 16205002 16210745 16210746 16210748 16210749 16153241 16153310 16153311 16203932 16193697 16193732 16193734 16193735 16193739 16195297 16193741 16193742 16193743 16192222 16195298 16193751 16193753 16189093 16193754 16193756 16192224 16193774 16193778 16195311 16207187 16207188 16207189 16207192 16190497 16118937 16190498 16173586 16143823 16143824 16198456 16198502 16203872 16203921 16192611 16192612 16196621 16192615 16192620 16173777 16173783 16173785 16173787 16173790 16173800 16098976 16098981 16125453 16125454 16125457 16125458 16125460 16125463 16193762 16193767 16193768 16193771 16169303 16169304 16198480 16204880 16204883 16204899 16177869 16178069 16156526 16156528 16156577 16189042 16192253 16192254 16192277 16192278 16192279 16192280 16192281 16192284 16192288 16192289 16192291 16200819 16203787 16203791 16210744 16207173 16207175 16207184 16207185 16195324 16195325 16199196 16195861 16177876 16207748 16200810 16200811 16211177 16211178 16211179 16211180 16167291 16198402 16198404 16198419 16198420 16199943 16199948 16199950 16195251 16173792 16173794 16173796 16173798 16176334 16177040 16177050 16202499 16202506 16180562 16200868 16199916 16199918 16188649 16196662 16146624 16146625 16189664 16189668 16189671 16189673 16189742 16180600 16180601 16178063 16173734 16173735 16173736 16173738 16187596 16187598 16184741 16199955 16199957 16087252 16087253 16199200 16207809 16207815 16195249 16204982 16204994 16213671 16213672 16071847 16071849 16071852 16071853 16071855 16071856 16071858 16203953 16203957 16203965 16206936 16206938 16206942 16206944 16206945 16194281 16194285 16211798 16048689 16164536 16167465 16203954 16203956 16203962 16203966 16203967 16183839 16204037 16190357 16190359 16167502 16195879 16195880 16209752 16194300 16194303 16210387 16211802 16213667 16164533 16164534 16216804 16216807 16210377 16203958 16203969 16204470 16204471 16190352 16190354 16190373 16190376 16190398 16208625 16209034 16209037 16216810 16216815 16220435 16207833 16207835 16203836 16143929 16143930 16143931 16177803 16212308 16199903 16189676 16189678 16189681 16189683 16189685 16189687 16189690 16189693 16189697 16189701 16189704 16189707 16189719 16189729 16214583 16214585 16214643 16167375 16167378 16167379 16216707 16216708 16121457 16208594 16208595 16209607 16199181 16199187 16199188 16199190 16216803 16216808 16216809 16216816 16167537 16167538 16155949 16155951 16155953 16155961 16198541 16199121 16199217 16211425 16211426 16216805 16209077 16209078 16218213 16218215 16218218 16218291 16210351 16214467 16214468 16206941 16206946 16206947 16207681 16207683 16208621 16208622 16209611 16209617 16208976 16208977 16208979 16198528 16198544 16167387 16167376 16220347 16220350 16220357 16220358 16220359 16220360 16220363 16220364 16207067 16207069 16207099 16207107 16214466 16211803 16202491 16202501 16202507 16202509 16202510 16202512 16202513 16202514 16203168 16223582 16161090 16190395 16214622 16214628 16214629 16209035 16185641 16187305 16216800 16216802 16216806 16216811 16216812 16216814 16216817 16213643 16207756 16207757 16167531 16167532 16167533 16211710 16211711 16211712 16190347 16190348 16208956 16208978 16136714 16136715 16136716 16216752 16216753 16207140 16098989 16169314 16169315 16209689 16209692 16218174 16218175 16218176 16218245 16218247 16219906 16218248 16217380 16217384 16217386 16217389 16217390 16217391 16217395 16217405 16217410 16217413 16204469 16212318 16211168 16167388 16216960 16216965 16218178 16218179 16218180 16218181 16218182 16220348 16220349 16220361 16220366 16192999 16210379 16155948 16167543 16167544 16155962 16189741 16189746 16189750 16189751 16214617 16214618 16214619 16214623 16209609 16209610 16198531 16216963 16213727 16205030 16205032 16190525 16190528 16190531 16190534 16162531 16102571 16172923 16216970 16187351 16206971 16206979 16206982 16206984 16206985 16206988 16206990 16206992 16207002 16207005 16207011 16207014 16207019 16207023 16207031 16207036 16207038 16207040 16207041 16195248 16217366 16160422 16160423 16160425 16190378 16148180 16146651 16146652 16212313 16212320 16212331 16212340 16212341 16212347 16208345 16146692 16146693 16167381 16211322 16188840 16188841 16188842 16188843 16188844 16188845 16188846 16218198 16218201 16209621 16219904 16219905 16218246 16219907 16220356 16221949 16187324 16189733 16189737 16189807 16214645 16198533 16188784 16216751 16209092 16216756 16216757 16209093 16213035 16203811 16203814 16203818 16203822 16203825 16203827 16203830 16203833 16162530 16176344 16176350 16205240 16205241 16205259 16205260 16205265 16205046 16209038 16209041 16211413 16211415 16163314 16221248 16169381 16221630 16221631 16187335 16187336 16187350 16187356 16187358 16187361 16187362 16187364 16187369 16207786 16207789 16207793 16209749 16040663 16040664 16102572 16176348 16176351 16176352 16207764 16207766 16205045 16214637 16214639 16210794 16210795 16210796 16210797 16210798 16210799 16210800 16210801 16210802 16210803 16210804 16210805 16210807 16210808 16210809 16210810 16210811 16210812 16210815 16210817 16210818 16219847 16202593 16202595 16212361 16212363 16203094 16203104 16203106 16189621 16189622 16189623 16189625 16189627 16189626 16216959 16212973 16204501 16204503 16204506 16213701 16213713 16210753 16210754 16176940 16218223 16218225 16218228 16218230 16218232 16213025 16213026 16213027 16217323 16217324 16218296 16207057 16207061 16207063 16207066 16203785 16203786 16207792 16207794 15638517 16200856 16200857 16221946 16221951 16221952 16221960 16219681 16192020 16190515 16211784 16148136 16148137 16176345 16176349 16210648 16213686 16213691 16205043 16205044 16195923 16195924 16195928 16195939 16195964 16195969 16195976 16195980 16195987 16196624 16196673 16196678 16198475 16215465 16215484 16215486 16169364 16169365 16169366 16210813 16210814 16210816 16211417 16207723 16207724 16207727 16207733 16207735 16207737 16207740 16207742 16216962 16209082 16209085 16209086 16209087 16209088 16209091 16190339 16190340 16195253 16153401 16155174 16153404 16160273 16202606 16212906 16212907 16212910 16212912 16212913 16212914 16212915 16212916 16212922 16219894 16219895 16219896 16221632 16221635 16223581 16189745 16221709 16219856 16216701 16196711 16196712 16196713 16196714 16196715 16196716 16196717 16213709 16213720 16219669 16219749 16215468 16211190 16211191 16211192 16211193 16207209 16207220 16207228 16220443 16220444 16220445 16220446 16220447 16220448 16220449 16220450 16218302 16218303 16222634 16219688 16222248 16109936 16109940 16218171 16218172 16219854 16221623 16221624 16221625 16205761 16205762 16205778 16205763 16205764 16205770 15955141 16221943 16221953 16176346 16176363 16209702 16209703 16195970 16195984 16195994 16196674 16196683 16196684 16212909 16212917 16212919 16212921 16212924 16212935 16216793 16216794 16214620 16214625 16003148 16222616 16183994 16183997 16217483 16217484 16215458 16218830 16218511 16219859 16219861 16221636 16194279 16194283 16194284 16194286 16194298 16194299 16153400 16153402 16153407 16153508 16153510 16153511 16153513 16153516 16194343 16211411 16190312 16190334 16222449 15638516 16219703 16219704 16217379 16217403 16063285 16063286 16214616 16214621 16214626 16219858 16219860 16219717 16219723 16219714 16219725 16219726 16217260 16219734 16219742 16219744 16109937 16158285 16214624 16207210 16207221 16207226 16207227 16216961 16216967 16216969 16224580 16224582 16224583 16224584 16224585 16224586 16224587 16224588 16224589 16225305 16153403 16153406 16153515 16176347 16176354 16219745 16223299 16195240 16216895 16216896 16213692 16189759 16214577 16214651 16216913 16216916 16216919 16209033 16195354 16209039 16209040 16209042 16209043 16216964 16216966 16218980 16218983 16211357 16222434 16222436 16208957 16208958 16208962 16208974 16219794 16219796 16219912 16217277 16221919 16221920 16221921 16221922 16209717 16209718 16209728 16209729 16209730 16209732 16218840 16214439 16211654 16211655 16211657 16211662 16227832 16216724 16216725 16219736 16219738 16219741 16219743 16222260 16222262 16222263 16223320 16223321 16210041 16210037 16210038 16205767 16210060 16210039 16205769 16210040 16205717 16222619 16222620 16222621 16222622 16222623 16222624 16222625 16225985 16225989 16221961 16221962 16221963 16204485 16222289 16222291 16222292 16221789 16221794 16223218 16223219 16221626 16221627 16223221 16223225 16225312 16219740 16225966 16225967 16225968 16211736 16211737 16211738 16222386 16222387 16211739 16211742 16222388 16211752 16222391 16211753 16222392 16211758 16222393 16222395 16220381 16220385 16220386 16220388 16213087 16213088 16222753 16222754 16222755 16196682 16216915 16216917 16217464 16217471 16220245 16220248 16220249 16220250 16220251 16220252 16220253 16220254 16220255 16220256 16220257 16220258 16220259 16220260 16220261 16220313 16227833 16227840 16224488 16224489 16224499 16228668 16225309 16218481 16218482 16188528 16188529 16188530 16188532 16188534 16188535 16188536 16188537 16188539 16188541 16188543 16188545 16210340 16210343 16188547 16188548 16188549 16219302 16219304 16209736 16209741 16208959 16208960 16208967 16208970 16208972 16208973 16221731 16221791 16221792 16225956 16205768 16217382 16223284 16223285 16185594 16221787 16221788 16212860 16212861 16212862 16212873 16212875 16212878 16212879 16212908 16217143 16173757 16173760 16225939 16220331 16231489 16231490 16231491 16231492 16173451 16173459 16173460 16173461 16173464 16173484 16173498 16173503 16220382 16220383 16220384 16220387 16109939 16109941 16223265 16223268 16225981 16219441 16219443 16219463 16189806 16173548 16173549 16173550 16173551 16173552 16173553 16173554 16173555 16173556 16173557 16225982 16225986 16225987 16225988 16225992 16222443 16222447 16202597 16202600 16202615 16202617 16221730 16221777 16212799 16212804 16223226 16223227 16220324 16225294 16225296 16222249 16173534 16209616 16208961 16208964 16208968 16208969 16208975 16229980 16230005 16222276 16227420 16228677 16228704 16220830 16220832 16220833 16220834 16224273 16222617 16222618 16216898 16216899 16220757 16220758 16220759 16216910 16222445 16226078 16226082 16227831 16230265 16227845 16225165 16230874 15638520 16222385 16211740 16211741 16222389 16222390 16211757 16211759 16208406 16208410 16214486 16214488 16214489 16214522 16214523 16214525 16214527 16214529 16214551 16214552 16214553 16214554 16216897 16222409 16203961 16222410 16222411 16222412 16222414 16222415 16222416 16222417 16214630 16199211 16199213 16199215 16222626 16209036 16221727 16221728 16221729 16222695 16222696 16222697 16222700 16222701 16222702 16222703 16222704 16222705 16222706 16222707 16222708 16222709 16222710 16213740 16222711 16192192 16222278 16200808 16219733 16219737 16219739 16219746 16219747 16219483 16219484 16215426 16213784 16217358 16217361 16164529 16225980 16225983 16226029 16226030 16226066 16226070 16208963 16230004 16220315 16172816 16223231 16216722 16220831 16220835 16220836 16220837 16222274 16230829 16230830 16232604 16203922 16222790 16222803 16210361 16210364 16210365 16210366 16210368 16208966 16221778 16221784 16221793 16226505 16226506 16227839 16223237 16223238 16223241 16223298 16228734 16228735 16223307 16228736 16223309 16228740 16223311 16228744 16223313 16228745 16223314 16228746 16222654 16222655 16222656 16222657 16222659 16222660 16222661 16222662 16222663 16222664 16222665 16222666 16222667 16222668 16121823 16121824 16121825 16225984 16226036 16226069 16218564 16218565 16218567 16213616 16213625 16213628 16219730 16216723 16219735 16219748 16228381 16228383 16230506 16228385 16230508 16230509 16221710 16222450 16231576 16231579 16219693 16219694 16219709 16219710 16227484 16227491 16227492 16227493 16227497 16227498 16227500 16227696 16227698 16227699 16227702 16227703 16222444 16225646 16221718 16216926 16216927 16227821 16230266 16222239 16226027 16226038 16226047 16226049 16221776 16221790 16216796 16216797 16216798 16216799 16190539 16218308 16226040 16226057 16222691 16222692 16222693 16222694 16221785 16221795 16226514 16227846 16228669 16228379 16230507 16217340 16217343 16225979 16225991 16226042 16226065 16199123 16199125 16199127 16199130 16199203 16199205 16199207 16199209 16127168 16127169 16127170 16213540 16213542 16213543 16213545 16213550 16213551 16213552 16213554 16218568 16218836 16218838 16218839 16218841 16224235 16224236 16224238 16224239 16224257 16224269 16225310 16195242 16222772 16222773 16222774 16222775 16230828 16236887 16236889 16214428 16214436 16219716 16223211 16223251 16223252 16219719 16223256 16223257 16219721 16219715 16223259 16212855 16212858 16212874 16212911 16226498 16226499 16226507 16232212 16210396 16210405 16167558 16220244 16220247 16220312 16230862 16221628 16227874 16230504 16230505 16228079 16228081 16228082 16217341 16210404 16210406 16210408 16210409 16210410 16227487 16227502 16227514 16227706 16222215 16222216 16222217 16222218 16222219 16222221 16222222 16222223 16198546 16199115 16199117 16199119 16230863 16226414 16226415 16228590 16120117 16236888 16236890 16218281 16204953 16204954 16204955 16204956 16204957 16204958 16204959 16204969 16204970 16204971 16204972 16204973 16204974 16204976 16219804 16219828 16223209 16223210 16223250 16223253 16223254 16223255 16223258 16223260 16210649 16210650 16210652 16210653 16210654 16210655 16210656 16210659 16210660 16210661 16210665 16210666 16230860 16230861 16230864 16230865 16167519 16223261 16223262 16233626 16233627 16233628 16239488 16239489 16239492 16225316 16220272 16220273 16220274 16220336 16220337 16224501 16224510 16230025 16230026 16218566 16218837 16209099 16209101 16209102 16209103 16209104 16211442 16211443 16211444 16213619 16213622 16222277 16081822 16221705 16227473 16227474 16227476 16227477 16227478 16227479 16227482 16227483 16227485 16227488 16227489 16227490 16227495 16227496 16227499 16227501 16227503 16227504 16227507 16227508 16227509 16227510 16227512 16227700 16227701 16227705 16227707 16227708 16236891 16236977 16228640 16228641 16228642 16228643 16228644 16227261 16227263 16227265 16227267 16227269 16227271 16227273 16227275 16227277 16227279 16227289 16227292 16227294 16227296 16227298 16227300 16227128 16223286 16228050 16228051 16228052 16228054 16228061 16228062 16228073 16228074 16228077 16228078 16224225 16224226 16224229 16224230 16224232 16224233 16224242 16224248 16218257 16087668 16087669 16087670 16087671 16205076 16208230 16208565 16208571 16208572 16208573 16208581 16208582 16208583 16236976 16219445 16219446 16219473 16219479 16227826 16233637 16220333 16231570 16231571 16231572 16231573 16231574 16231575 16231577 16231578 16231580 16240958 16048721 16204549 16204553 16010058 16010059 16010060 16010061 16010062 16010063 16010064 16208413 16208416 16231495 16231496 16231497 16231498 16231499 16231500 16230550 16230823 16230825 16236884 16223663 16226519 16226520 16226522 16226526 16226527 16226528 16226529 16226530 16226531 16226533 16226574 16226578 16226582 16230037 16230039 16226500 16226501 16226502 16226503 16226508 16226509 16226513 16217332 16222651 16234058 16234059 16236908 16236909 16236914 16236916 16234112 16236920 16226422 16226427 16231015 16220288 16208555 16208557 16208584 16220317 16222226 16219450 16219461 16219464 16219468 16228044 16228053 16228063 16228065 16228068 16228071 16231049 16231050 16231052 16231053 16205064 16205068 16205074 16205082 16199866 16222224 16209615 16236892 16219802 16219809 16219820 16219822 16219827 16219829 16219830 16230036 16230038 16213635 16240314 16228595 16224216 16224246 16227879 16227880 16235888 16235889 16235894 16232414 16232415 16232416 16232417 16231017 16231018 16231019 16222446 16230362 16216918 16222448 16231054 16229937 16226504 16226510 16226511 16226512 16221629 16233485 16235908 16234035 16234036 15998629 16234739 16222796 16239119 16239124 16239125 16236937 16233647 16233652 16233653 16236903 16236904 16222627 16222628 16228094 16210662 16210663 16195849 16189082 16227828 16234798 16234801 16230502 16230503 16230542 16230552 16230553 16230554 16230555 16230556 16240759 16235907 16235909 16153221 16153222 16087666 16087667 16087672 16087673 16163216 16077492 16176255 16235932 16230833 16230835 16230836 16230873 16230837 16230838 16230839 16230844 16230846 16240752 16234110 16236917 16221704 16221706 16214555 16214564 16227475 16227480 16227481 16227486 16227494 16227505 16227506 16227511 16227513 16240315 16227695 16227704 16227709 16227710 16227712 16208231 16236938 16230547 16230549 16230826 16241598 16241603 16241604 16219807 16219815 16219825 16236984 16234786 16228089 16228090 16228093 16228095 16210658 16223703 16228804 16235971 16235972 16213618 16213621 16213627 16213630 16213632 16213633 16218204 16238044 16238060 16238088 16218206 16218207 16218210 16218211 16224212 16224213 16224218 16224219 16224221 16224222 16234795 16234799 16234800 16235885 16225970 16235910 16195244 16239147 16239149 16239292 16239293 16239295 16239296 16239297 16239298 16239299 16239300 16222797 16213808 16234742 16234109 16234111 16236918 16236919 16239106 16239108 16240999 16239110 16239112 16239114 16239116 16218311 16236936 16146645 16146646 16146647 16230358 16230359 16230360 16224511 16224512 16224513 16224515 16219806 16219811 16219817 16219831 16219832 16232612 16228088 16228091 16226517 16226518 16226521 16226523 16226524 16226525 16226576 16226579 16226580 16212970 16234057 16238070 16238071 16238087 16244934 16244158 16232382 16236849 16236850 16236851 16236852 16236853 16230561 16230564 16230765 16230766 16230776 16230777 16230779 16230780 16230782 16213765 16213766 16213776 15998618 15998620 15998621 15998622 15998624 15998625 15998626 15998627 15998628 15998630 16240546 16230840 16230842 16230843 16230845 16230847 16230849 16230850 16230851 16230852 16230855 16230856 16230857 16230858 16229947 16229949 16230557 16230559 16240657 16241536 16231026 16231027 16231028 16231030 16231032 16231034 16231036 16231042 16231044 16231047 16231048 16236935 16236939 16236942 16241601 16230558 16230562 16240655 16241643 16241644 16230560 16230565 16209746 16228092 16228096 16232621 16232622 16228109 16244296 16241532 16241533 16232384 16241534 16241535 16240659 16240588 16234039 16234041 16234042 16234043 16236940 16236941 16173531 16173533 16230827 16198522 16198524 16198526 16210855 16210856 16208991 16215472 16232613 16232614 16240672 16232578 16232366 16232367 16241646 16241648 16241652 16241654 16241655 16245191 16245192 16245194 16245196 16153223 16241600 16187345 16209647 16240997 16222394 16204998 16222396 16222397 16222398 16222399 16222400 16222402 16235181 16235183 16235184 16235185 16235186 16235188 16235189 16235190 16235191 16235193 16235864 16235865 16235882 16229908 16229911 16229936 16234047 16234048 16234049 16234050 16234051 16234052 16234053 16234054 16234116 16234117 16211685 16211686 16211687 16211688 16211689 16241651 16245193 16245195 16245197 16227910 16230268 16241236 16220841 16221694 16221708 16245153 16235182 16239498 16239499 16241488 16241490 16241492 16241495 16226083 16225172 16225173 16225174 16225175 16225176 16225177 16225178 16225180 16225182 16225184 16244643 16244644 16244656 16244187 16244189 16226007 16226008 16236859 16236860 16244967 16219813 16219818 16219823 16219824 16219826 16219833 16219834 16219835 16219836 16219843 16240960 16240961 16240962 16240965 16240966 16240967 16222658 16222776 16222777 16222778 16222779 16222780 16222781 16222782 16222783 16222784 16222785 16222786 16227905 16227906 16230872 16244948 16176238 16087677 16222406 16222407 16251827 16234775 16234783 16240959 16240968 16228783 16230029 16220812 16220813 16220814 16220815 16220816 16219351 16219352 16219353 16219354 16223642 16223643 16224207 16224209 16224210 16224247 16230539 16224256 16224260 16224262 16233641 16233642 16245476 16245884 16245155 16245156 16245163 16245164 16245165 16245166 16245411 16226426 16231014 16209671 16230796 16230797 16230799 16230816 16230819 16230820 16241627 16241630 16240382 16226515 16213747 16213753 16213775 16213777 16231022 16231023 16231024 16231025 16231029 16231031 16231038 16231040 16231043 16063193 16241547 16121462 16222718 16222719 16222721 16240198 16240199 16240319 16228778 16228781 16172504 16245423 16245426 16245427 16245430 16245431 16241395 16241396 16241397 16241398 16241399 16241400 16241401 16241402 16241403 16230563 16244661 16244662 16244664 16222629 16222630 16222631 16224522 16087674 16168879 16215357 16198492 16198496 16198535 16198539 16208992 16236875 16236877 16236882 16236883 16222225 16222733 16222734 16222735 16222736 16222737 16222738 16222740 16222742 16222744 16222746 16241628 16241632 16240679 16240681 16225960 16225962 16240745 16232353 16248947 16248948 16248949 16248950 16230273 16230274 16230275 16230276 16231033 16231035 16231037 16231039 16231041 16234056 16234060 16236907 16236910 16236911 16236913 16236915 16240461 16240463 16240465 16240467 16240468 16241226 16234113 16192313 16192317 16192319 16198378 16198380 16195828 16241596 16251823 16240963 16240964 16223664 16223665 16223666 16223667 16235939 16235940 16235941 16235943 16235944 16235945 16235946 16235947 16235948 16235967 16230793 16230801 16230803 16230812 16230813 16228777 16241631 16241638 16241640 16233483 16238031 16238045 16238055 16238075 16238056 16238057 16238061 16238079 16238080 16238089 16238090 16250336 16250337 16250341 16246327 16246330 16246332 16246337 16246340 16246344 16250342 16232390 16247346 16247350 16230785 16230786 16230789 16230790 16230869 16236923 16236924 16236925 16240746 16240747 16240748 16240749 16240750 16230841 16240751 16222652 16177853 16155265 16245417 16230815 16231020 16244952 16241642 16240398 16240402 16240404 16241493 16241494 16241496 16240753 16240754 16236944 16236945 16236946 16236947 16245424 16245425 16219448 16219452 16219454 16219455 16219466 16219470 16219471 16219482 16233608 16223620 16223622 16223624 16223626 16223628 16223629 16223630 16223631 16223632 16231021 16248729 16248730 16248736 16239485 16287936 16236854 16236855 16236856 16236858 16236861 16236863 16252574 16252580 16236885 16236886 16198505 16198507 16198509 16198511 16246408 16246409 16241489 16241491 16245428 16245429 16248727 16248728 16240956 16241257 16244907 16244910 16244913 16244953 16244955 16220838 16220840 16221692 16221695 16065384 16241388 16241389 16241392 16248963 16250317 16250318 16250319 16250320 16250321 16250322 16250323 16250324 16250325 16235942 16235968 16230818 16230822 16230824 16226532 16241615 16241617 16241619 16241620 16241622 16241633 16241634 16241635 16241639 16241641 16238052 16244141 16244142 16252991 16223585 16223621 16223627 16223634 16223635 16223636 16223637 16240755 16240756 16219000 16219002 16219004 16219007 16219016 16219280 16219295 16219296 16219298 16219342 16250388 16250389 16253573 16241537 16247357 16247358 16241539 16241540 16250391 16250392 16250393 16244181 16230511 16230512 16230513 16230514 16230515 16230516 16230517 16230518 16230519 16230520 16230521 16230522 16230523 16230524 16230525 16230526 16230527 16230528 16230529 16230530 16230531 16230532 16230533 16230534 16230535 16230536 16230537 16230538 16240957 16240620 16241259 16241260 16219700 16244898 16244900 16244958 16244960 16244962 16234044 16234045 16241587 16241001 16241208 16241589 16241590 16240472 16241880 16240475 16244989 16244991 16241883 16192324 16192328 16241605 16251829 16230363 16230366 16240595 16244208 16244599 16244601 16244676 16244678 16244680 16244682 16244684 16244686 16244690 16244693 16233801 16233802 16226075 16240757 16240758 16241233 16241235 16230271 16240322 16240200 16240163 16240165 16240175 16240177 16240183 16241355 16240007 16222431 16231016 16250351 16241404 16241405 16241406 16241407 16237048 16237050 16237052 16237053 16237054 16237055 16237056 16237057 16237059 16237060 16237061 16237063 16231051 16245101 16235995 16235996 16235997 16235998 16235999 16253204 16252620 16252621 16252623 16230783 16230787 16230788 16230791 16246241 16234038 16234040 16234046 16241541 16241385 16241386 16241387 16248041 16248043 16248045 16248047 16252562 16241391 16241624 16241629 16241636 16241637 16250355 16250356 16247331 16247334 16247342 16247343 16234803 16253202 16253205 16239481 16239482 16239483 16241542 16241543 16241892 16241544 16252571 16252572 16252573 16252575 16252576 16252577 16252578 16252579 16251921 16251930 16240675 16240677 16224266 16216981 16253203 16253151 16241234 16230784 16230867 16230868 16248759 16253168 16253169 16244545 16244548 16244650 16243836 16243838 16243840 16243845 16241394 16241475 16241481 16241498 16241512 16236926 16241882 16219765 16219767 16219769 16219771 16219775 16219779 16219782 16236874 16236876 16236879 16236881 16228281 16228290 16228295 16228393 16228394 16228395 16228403 16228405 16228406 16228408 16228460 16254130 16252630 16254142 16252980 16241538 16253154 16250390 16253082 16253083 16253144 16253150 16218293 16253185 16255058 16246272 16241896 16241897 16254095 16254097 16254100 16254106 16236870 16236871 16236872 16222636 16222638 16222642 16120076 16250451 16250454 16241449 16241451 16241452 16241456 16247329 16247333 16246274 16168689 16222802 16246277 16252560 16253231 16253232 16244609 16244610 16244611 16244613 16223623 16223625 16222413 16184105 16184107 16188490 16184123 16184144 16184793 16184795 16184797 16184809 16185491 16185492 16185495 16185498 16185501 16185522 16185524 16185532 16185535 16185539 16185541 16216908 16234784 16250452 16244524 16244538 16219457 16219459 16219477 16219481 16218235 16255046 16245981 16245987 16240070 15917861 15917863 16241546 16241548 16241549 16215462 16215487 16260247 16260281 16250453 16247284 16247288 16247289 16247291 16247294 16247295 16234118 16234119 16234120 16234121 16253016 16247356 16254431 16253156 16250447 16250448 16250449 16250450 16256455 16256457 16256458 16230794 16230795 16230800 16230804 16230805 16230806 16230811 16230814 16230817 16255043 16251891 16241898 16250458 16250459 16250461 16250462 16250463 16250464 16250467 16250466 16250470 16250473 16250476 16226445 16226446 16187758 16187759 16187764 16251934 16251937 16223701 16223702 16228291 16228293 16228294 16228369 16228392 16228404 16228407 16228445 16228446 16228454 16228468 16228469 16240380 16233803 16241545 16243825 16241887 16241384 16241889 16241890 16241891 16243827 16243828 16244118 16244119 16245304 16260262 16209009 16209021 16209024 16209026 16245098 16227801 16227802 16227803 16227804 16227805 16227806 16227807 16227808 16244932 16246328 16244936 16244942 16250343 16252990 16232374 16232380 16225186 16255737 16255744 16254485 16254486 16255716 16254495 16241513 16241390 16241393 16222404 16222405 16227531 16227533 16227535 16227537 16227539 16227541 16228042 16228043 16240396 16240400 16240142 16241881 16241884 16244117 16244120 16215470 16215488 16215492 16215495 16215497 16245089 16245091 16251925 16251926 16251933 16228278 16228279 16228280 16228282 16228292 16228459 16228472 16211727 16211728 16230267 16243815 16243816 16253155 16254432 16255053 16245983 16260293 16260294 16260295 16256456 16252585 16253372 16251901 16251903 16261516 16241582 16230017 16241888 16241894 16202735 16121461 16260251 16260273 16241531 16248321 16253564 16228040 16258427 16248953 16248954 16266064 16240765 16240764 16240952 16241359 16241360 16241362 16241364 16241365 16241457 16241459 16241460 16241463 16241464 16241465 16241469 16241470 16241476 16241478 16241482 16241483 16241484 16241885 16241886 16260164 16260169 16244523 16244530 16244532 16244533 16244534 16244535 16241447 16241450 16241453 16250326 16250327 16250328 16250329 16250330 16250331 16250338 16250339 16250340 16232351 16232356 16232358 16232372 16218708 16230384 16253578 16253579 16253584 16253604 16254093 16262504 16254098 16254105 16250334 16245074 16241583 16241584 16241585 16258434 16252100 16264716 16240766 16240767 16240953 16240954 16241514 16250332 16250333 16250335 16259570 16259571 16260156 16260210 16230809 16230810 16240684 16240702 16250446 16266208 16258429 16258433 16264708 16260296 16246021 16246231 16246238 16246240 16246246 16246247 16188976 16251814 16251816 16246304 16246314 16246319 16246322 16246370 16246371 16246372 16246373 16260155 16260217 16260219 16245062 16245076 16257811 16258374 16258378 16258379 16258381 16258383 16161372 16251872 16251875 16223633 16223638 16259504 16259505 16259506 16224537 16224538 16224541 16224542 16224544 16224546 16224549 16224550 16224552 16240628 16240632 16241356 16241357 16199931 16199932 16240988 16133222 16133226 16133227 16240989 16240990 16240991 16240993 16240994 16244738 16244750 16256395 16256396 16256405 16256410 16250455 16250456 16250469 16250471 16250468 16250472 16250474 16250477 16250478 16260193 16223687 16254530 16254531 16254532 16254552 16253501 16253503 16253507 16253510 16254200 16254208 16254211 16254213 16254215 16254218 16254419 16262503 16262507 16212306 16210860 16215494 16132427 16224600 16224602 16224611 16224612 16224614 16224615 16224632 16254545 16254546 16254547 16254549 16250359 16250360 16250363 16250364 16250371 16250372 16250373 16250374 16250375 16230381 16219278 16219293 16219294 16219297 16219325 16219331 16219336 16219344 16236906 16254496 16256392 16256394 16227697 16256399 16256400 16256402 16256403 16256404 16256407 16256408 16256409 16256411 16256413 16256414 16256415 16202699 16202708 16189739 16253571 16253572 16224613 16224617 16254526 16254540 16253500 16253504 16253505 16253506 16253508 16253509 16260355 16255075 16220761 16224214 16224227 16224254 16224255 16241581 16225181 16266117 16266119 16266121 16266122 16266123 16266124 16224554 16222632 16240458 16259468 16259471 16244752 16244753 16244754 15945379 15945380 16234748 16234759 16234760 16256498 16256500 16256502 16182177 16256391 16256393 16256397 16256398 16256401 16256406 16261578 16261594 16245162 16245412 16230353 16230354 16230356 16230357 16165101 16259572 16259573 16259574 16259575 16260163 16260170 16260171 16260175 16260186 16260194 16240591 16245078 16230802 16230821 16244945 16260286 16244946 16254534 16254542 16254550 16254551 16254553 16222463 16222470 16222471 16230382 16264558 16219475 16263548 16221637 16234114 16259514 16269304 16241895 16258474 16258475 16258477 16258478 16258479 16258481 16258482 16258484 16253300 16253308 16241588 16250482 16250483 16250484 16250485 16250486 16240470 16250487 16253312 16253314 16253316 16255698 16253319 16260329 16244984 16253323 16244986 16260336 16253326 16241227 16255703 16251963 16255704 16255062 16255063 16270526 16270530 16254537 16254543 16254548 16240648 16240995 16240996 16252981 16234754 16234755 16227126 16256513 16257598 16241893 16250460 16250465 16223688 16250365 16228258 16269290 16269291 16269292 16245994 16245995 16270124 16205065 16205069 16205072 16205081 16205083 16205084 16205205 16205206 16267718 16267719 16267720 16267721 16267723 16267724 16250357 16264648 16263539 16263542 16263547 16266118 16266125 16266126 16266127 16266128 16252096 16252097 16252103 16252104 16252105 16252106 16263263 16263264 16261548 16269986 16269987 16269989 16269990 16271391 16271392 16240950 16240951 16253367 16253214 16270508 16270510 16270972 16253376 16253241 16253246 16253248 16253359 16254527 16254533 16254535 16254536 16254541 16254554 16254555 16254556 16254557 16266919 16266922 16263550 16247191 16247200 16247202 16247203 16247204 16247205 16247206 16247208 16247210 16247211 16247212 16071851 16202703 16251876 16251878 16252538 16252539 16261480 16252544 16261482 16261483 16261490 16263538 16266929 16266120 16274785 16271377 16271378 16230510 16271390 16253173 16213793 16252526 16252527 16252528 16252533 16252536 16246311 16246312 16246315 16246368 16246369 16246374 16264640 16264646 16264647 16243817 16243819 16269988 16272551 16272552 16272553 16272554 16269306 16271398 16264712 16269309 16264714 16264541 16264542 16264543 16264544 16240761 16240762 16241358 16241366 16234762 16270573 16273861 16236905 16264668 16274786 16271382 16241657 16241659 16241879 16264588 16258414 16257609 16257612 16250475 16259553 16267716 16262530 16262531 16262532 16253375 16253378 16253483 16253484 16253485 16253536 16267729 16253546 16253547 16253555 16267733 16267734 16267736 16267737 16253239 16253242 16253355 16251874 16252147 16241455 16248065 16250606 16251813 16269285 16271011 16271376 16246019 16246020 16246232 16246233 16241467 16241468 16241474 16241477 16241479 16241480 16241485 16241500 16241501 16241502 16241503 16241504 16241505 16241507 16241508 16241510 16241511 16241527 16241528 16246294 16246313 16246320 16246321 16236929 16236932 16237035 16237038 16237042 16237043 16253238 16253240 16253349 16254528 16254529 16252145 16251881 16262566 16262568 16247215 16247216 16247217 16247218 16247219 16247220 16247221 16247222 16247236 16247237 16247238 16247239 16247254 16247255 16247256 16247257 16247258 16247259 16247260 16247262 16247263 16247264 16240638 16240639 16240640 16240641 16240642 16240643 16240644 16240645 16240646 16240647 16266093 16266094 16266095 16266098 16266102 16266103 16266105 16273854 16273856 16273859 16276513 16264540 16245984 16246001 16274795 16275548 16275550 16275551 16274796 16274798 16275558 16239312 16254199 16262533 16262534 16253548 16267743 16270892 16270893 16273518 16266935 16273520 16264451 16269282 16269283 16274762 16274776 16274778 16274779 16251905 16271400 16240948 16240955 16241361 16241363 16260207 16260209 16260211 16211437 16211438 16251877 16276474 16276475 16248958 16204557 16238881 16270930 16244568 16244569 16269268 16264593 16260333 16264615 16264617 16264619 16253327 16239127 16259535 16259560 16259562 16259564 16223686 16241237 16241238 16241239 16241240 16241241 16241367 16241368 16241369 16241370 16241371 16241372 16241373 16241374 16241375 16241376 16241377 16241378 16241379 16241380 16241381 16241382 16241383 16241418 16241420 16241422 16241423 16241424 16241425 16241426 16241427 16241428 16241429 16241431 16241432 16241433 16241434 16241435 16241436 16241437 16241438 16241439 16241440 16241441 16241442 16241443 16270575 16273851 16266130 16266134 16266139 16266140 16266141 16266142 16266143 16272590 16272591 16258455 16246005 16246006 16238847 16270929 16270931 16256477 16256495 16257603 16257631 16257633 16276497 16276501 16276503 16263416 16263419 16263420 16259537 16259538 16245953 16245954 16245955 16245956 16245957 16245958 16245959 16245960 16245961 16245962 16245963 16245964 16245965 16245966 16245967 16245968 16245969 16245970 16245971 16245972 16245973 16245975 16245976 16245977 16245979 16270413 16270416 16270419 16270422 16251871 16251873 16252118 16246003 16273863 16273864 16273866 16273868 16273870 16273872 16273876 16273877 16273880 16270948 16266156 16266158 16266159 16266162 16266908 16218926 16218927 16218935 16218936 16218953 16218959 16218960 16218961 16218962 16097589 16276495 16276499 16230807 16230808 16251879 16251880 16268452 16268453 16270880 16270881 16266921 16280867 16270882 16245459 16266944 16266946 16275575 16276622 16261588 16275641 16275643 16276807 16262506 16276808 16276809 16259536 16256507 16276505 16236951 16236952 16237016 16237017 16237019 16237022 16237024 16237025 16237026 16246293 16246295 16246301 16246302 16251907 16251908 16240403 16240405 16240407 16271432 16271437 16271447 16271452 16271453 16270891 16263557 16273516 16273517 16270894 16270895 16266936 16266129 16266131 16266132 16266133 16266135 16266136 16266137 16266138 16266144 16266145 16266146 16266148 16282206 16282208 16274777 16271005 16271006 16271010 16240760 16254194 16238979 16254196 16238980 16198519 16198537 16210869 16215496 16257657 16257658 16257659 16257660 16257690 16257692 16257693 16257694 16257695 16257696 16257810 16246413 16246442 16246443 16247185 16247213 16241448 16244140 16244143 16273770 16282211 16282212 16282213 16269991 16268470 16268472 16268509 16268510 16268511 16268512 16256412 16256416 16256417 16212298 16212299 16241458 16241466 16241486 16241487 16241497 16270964 16269330 16244891 16270995 16260285 16282202 16166753 16261508 16261509 16261519 16268448 16268450 16276562 16276564 16230019 16219272 16219274 16219276 16219338 16220729 16281607 16220754 16273840 16270965 16248957 16263400 16266157 16234749 16240076 16254198 16270919 16270924 16227711 16274699 16274701 16274702 16274703 16274704 16274705 16274706 16274707 16274708 16274709 16274710 16274711 16274712 16274713 16274714 16274716 16274717 16274718 16274719 16272529 16269446 16276525 16269269 16269270 16269271 16217265 16217267 16217273 16273256 16208362 16236928 16236930 16236931 16236933 16237036 16237037 16237039 16237041 16237046 16237047 16241530 16230364 16223689 16224302 16224303 16224304 16253487 16253535 16267735 16273453 16273454 16273455 16273456 16273457 16241444 16241445 16259473 16259474 16283443 16283444 16283445 16283446 16268444 16268446 16268455 16268457 16268459 16268461 16268463 16276452 16276455 16268559 16268560 16268561 16268564 16273746 16273747 16273748 16273749 16273750 16273765 16270871 16220727 16281609 16220731 16220733 16220735 16220737 16220739 16220741 16220743 16220745 16220747 16220748 16220750 16281610 16281611 16266938 16270988 16277497 16274760 16274761 16274763 16282204 16282205 16270900 16270903 16270904 16246016 16246234 16281631 16281632 16276806 16270579 16241462 16237018 16237020 16274733 16237027 16237029 16237030 16237031 16237032 16222418 16222419 16204481 16246292 16246296 16246303 16283879 16283885 16253486 16267730 16241446 16269332 16283439 16283440 16283441 16246410 16264579 16240695 16246421 16246422 16246423 16246424 16246425 16246426 16247186 16247187 16247188 16247189 16247190 16247201 16250358 16250361 16250362 16250366 16250367 16250368 16250369 16250377 16241454 16250488 16266211 16287906 16281648 16281649 16276573 16280866 16283863 16283864 16259493 16244145 16282210 16276510 16271009 16286888 16285029 16285034 16286889 16285035 16244723 16245487 16283461 16283942 16247298 16247299 16247300 16247307 16247308 16281613 16273528 16273531 16273534 16281614 16273458 16283868 16277436 16277438 16277442 16277447 16277450 16273837 16273843 16283591 16283593 16286388 16286389 16282220 16282221 16282222 16263434 16241461 16241526 16273265 16270949 16250376 16241499 16241506 16241509 16236000 16236001 16236004 16236005 16236009 16236010 16236011 16241263 16236756 16236757 16236758 16236759 16236761 16236762 16236835 16236836 16236837 16236838 16236839 16236843 16236845 16236846 16236847 16273742 14791297 16187368 16255077 16255192 16255193 16255209 16255586 16256453 16256454 16293827 16198433 16204556 16270977 16231045 16231046 16253206 16253207 16267662 16267663 16267665 16267700 16268432 16283842 16283843 16283844 16283845 16283846 16283847 16283848 16283849 16283850 16283852 16283853 16283855 16273467 16273469 16273471 16273473 16273478 16273481 16273483 16273485 16273488 16273490 16273492 16273494 16273496 16273498 16273500 16273502 16283851 16283854 16283856 16283857 16283858 16283859 16283860 16246015 16246242 16246244 16246258 16246260 16166772 16166779 16166787 16062043 16062044 16227927 16230269 16227942 16227946 16230270 16227948 16227949 16227950 16227952 16227954 16055415 16218937 16218938 16218939 16218946 16218947 16218948 16218963 16218964 16218965 16218967 16218968 16218969 16218972 16218973 16218975 16275562 16275567 16275569 16275574 15944345 15944346 16273505 16273506 16273507 16273509 16296149 16296150 16296151 16296152 16296153 16234750 16234764 16234770 16234772 16234774 16263402 16263403 16263404 16114625 16257607 16257649 16257651 16234037 16286390 16286391 16286392 16286394 16274802 16274804 16274807 16274809 16274811 16277515 16277516 16277519 16277538 16277539 16277540 16277541 16277551 16177001 16177796 16177798 16283596 16283457 16283458 16253216 16253219 16282216 16236862 16253220 16276470 16276471 16220839 16221693 16221696 16221697 16221698 16221699 16221700 16221702 16221707 16286846 16286847 16269340 16281630 16281633 16281634 16274698 16286367 16286368 16286369 16286370 16286371 16286373 16286374 16286375 16286376 16286377 16286378 16286379 16286380 16286381 16286382 16286384 16282231 16282232 16286385 16285133 16285134 16262505 16285140 16285141 16244970 16253310 16260326 16282217 16260330 16282218 16282219 16276527 16276530 16264621 16264625 16290620 16276532 16260334 16285044 16276534 16264509 16139598 16286852 16290678 16286853 16286854 16290596 16268526 16268570 16268572 16268574 16268576 16269241 16269242 16226448 16239368 16212300 16218976 16218977 16164500 16254424 16295818 16247228 16248051 16248053 16164535 16287839 16287840 16287841 16287842 16287844 16287845 16287846 16287847 16287849 16287850 16287851 16287852 16287853 16288322 16288328 16288331 16288333 16288335 16288343 16288344 16288345 16288347 16273272 16273273 16273274 16287854 16287855 16222220 16283887 16283888 16085348 16246379 16246380 16246381 16263496 16263498 16263499 16263501 16263502 16263506 16226431 16289360 16288267 16241515 16287932 16287938 16287939 16288268 16267712 16267713 16246384 16246385 16246386 16293835 16246391 16246392 16246393 16246394 16246395 16246405 16246407 16237044 16237045 16267714 16283878 16283880 16283881 16283883 16283886 16138149 16245909 16283464 16283465 16283466 16283467 16283468 16283469 16283470 16283471 16283820 16283475 16283477 16283479 16283481 16283483 16283485 16283486 16283487 16283488 16283489 16283490 16283491 16283493 16283822 16283823 16283824 16283827 16283828 16241529 16288367 16288379 16288383 16288393 16288399 16293855 16240196 16240197 16293856 16293857 16293858 16294522 16296885 16223685 16224301 16224318 16224320 16224321 16224322 16224323 16224324 16224328 16224329 16164429 16164430 16164431 16164432 16164433 16164434 16164435 16164436 16164437 16164438 16267742 16220408 16220409 16220410 16220411 16220412 16220413 16220414 16220415 16220416 16220417 16220418 16220419 16220420 16220421 16220422 16220423 16220424 16220425 16220426 16273463 16273464 16230798 16269274 16269275 16211720 16211722 16268668 16269276 16269277 16260359 16260361 16258492 16258493 16258496 16258497 16258499 16258523 16263268 16263269 16263272 16263274 16258524 16258525 16263275 16263376 16263377 16260362 16260364 16259365 16259366 16261452 16261453 16261455 16261456 16261458 16261459 16277451 16277453 16277457 16277462 16277466 16277467 16277469 16277473 16277477 16229906 16240426 16240432 16240435 16276543 16231055 16231056 16226084 16226085 16268537 16268540 16270945 16283442 16286851 16290595 16281667 16285807 16281668 16283874 16290708 16290709 16290710 16288269 16288270 16288271 16266150 16294527 16240687 16240690 16240691 16240700 16270140 16280831 16280832 16280833 16280834 16280835 16280836 16280837 16280838 16280839 16280840 16280841 16280842 16280843 16280844 16280845 16280848 16280852 16280853 16280855 16280856 16280858 16280859 16280860 16280861 16236002 16236003 16236006 16236007 16236008 16236763 16236840 16236842 16236844 16236848 16246414 16246415 16246416 16246417 16246418 16246419 16247261 16247265 16259583 16263379 16259376 16261472 16261473 16261476 16263381 16259378 16263383 16263385 16263386 16259379 16250370 16281621 16281622 16281623 16281624 16281625 16281626 16281627 16281628 16281629 16281646 16283597 16295796 16283447 16295797 16283448 16295802 16295803 16283449 16283450 16283451 16295806 16283452 16283453 16295809 16295810 16295815 16296900 16296902 16296926 16296929 16296939 16296940 16296941 16296942 16296943 16296944 16296945 16272607 16295760 16295761 16295762 16295763 16293644 16293645 16293646 16295764 16293647 16295765 16295767 16293648 16295768 16295769 16295770 16295771 16295772 16295773 16296946 16296947 16296948 16296949 16296950 16296951 16296952 16296953 16296954 16296955 16296956 16296957 16296958 16296959 16296960 16296961 16296962 16236878 16236880 16269251 16269252 16269253 16269254 16269255 16269256 16286855 16286856 16296983 16296984 16286857 16286858 16296985 16296986 16286859 16286860 16286861 16276450 16276451 16276453 16276454 16276456 16276457 16276458 16276459 16244939 16286409 16287828 16287829 16287830 16287832 16290670 16298784 16287856 16298786 16290671 16287857 16285021 16285804 16290672 16288280 16288281 16264699 16289192 16276462 16276465 16276466 16276467 16288300 16288302 16288309 16288314 16288315 16286441 16286442 16286443 16286444 16286545 16286546 16286547 16286548 16286549 16286550 16286551 16286552 16286553 16286554 16286555 16286556 16286557 16286558 16270444 16270445 16270446 16270447 16270448 16270449 16259387 16259405 16259407 16259409 16259412 16259415 16259419 16259423 16259428 16263509 16263510 16286826 16286827 16286828 16286829 16286830 16286831 16286832 16286833 16286834 16282214 16282215 16277481 16270969 16255033 16290711 16290712 16290713 16290714 16290715 16290716 16290717 16290718 16290719 16287917 16287835 16287836 16286848 16286849 16298030 16298033 16298034 16280862 16280863 16287918 16287919 16287920 16289241 16289242 16289243 16289245 16289246 16289247 16289248 16289249 16289250 16289251 16289252 16289253 16289254 16289255 16289256 16269279 16276558 16276559 16270385 16270386 16270387 16270388 16270390 16270391 16270400 16270405 16270409 16286395 16286396 16286397 16286398 16286399 16286400 16286401 16286790 16286791 16286792 16286793 16286794 16286816 16286817 16262404 16262405 16262406 16262407 16262409 16262411 16286404 16283454 16286405 16286406 16286407 16283455 16286408 16283456 16240084 16240088 16240106 16224249 16224250 16240108 16240110 16240112 16275629 16288317 16288318 16283830 16283831 16283832 16269280 16273848 16283861 16283862 16296195 16289193 16287921 16293836 16286835 16286836 16287922 16286837 16283865 16287011 16287014 16287015 16287016 16287017 16287018 16288264 16288265 16288266 16290696 16280869 16280870 16280871 16280872 16280873 16280875 16281604 16281605 16281606 16219340 16297971 16297974 16296201 16286838 16286839 16286840 16289203 16287923 16289204 16289205 16287924 16289210 16283866 16283867 16281608 16285047 16285051 16285053 16285055 16285056 16285057 16285058 16285060 16285062 16285064 16285065 16285068 16285069 16285070 16285071 16285073 16285074 16285067 16285119 16296204 16296207 16296208 16245462 16245466 16288320 16289234 16289236 16277487 16277488 16281612 16286841 16287925 16287926 16283869 16289198 16289199 16286842 16287927 16296210 16296211 16293838 16293839 16296214 16293840 16286843 16286844 16253176 16216884 16290679 16290680 16290681 16290682 16290683 16290684 16290685 16290686 16290687 16290688 16290689 16290690 16290691 16290692 16296110 16296111 16296112 16296113 16296114 16296115 16296117 16296118 16296119 16296120 16296121 16296122 16288401 16288402 16276623 16276624 16276625 16276626 16276627 16277424 16277425 16277426 16277427 16277428 16277430 16277432 16277434 16277440 16277444 16277449 16286351 16286352 16286353 16286354 16286355 16286356 16286357 16286358 16286360 16286361 16286362 16286363 16286364 16286365 16287858 16287859 16287860 16287861 16287863 16287864 16287865 16286862 16287866 16287867 16287868 16287869 16287870 16287871 16288349 16288350 16288352 16288353 16288354 16288355 16288356 16288357 16288358 16288359 16288360 16288361 16288362 16269284 16282203 16282207 16296155 16296157 16283590 16290669 16297976 16297977 16297980 16283592 16286387 16287913 16288403 16287914 16288558 16287915 16287916 16283594 16276547 16276548 16285030 16285032 16285033 16285127 16290643 16290667 16283871 16283872 16293841 16293842 16293843 16281650 16290593 16281615 16281616 16281617 16281618 16281619 16281620 16287928 16287929 16287930 16293844 16287931 16293845 16293847 16270953 16270954 16270955 16281654 16270956 16293846 16218297 16259516 16289237 16286850 16288263 16298787 16297987 16297988 16289238 16294528 16288274 16288275 16297993 16297997 16297998 16295530 16295529 16289239 16286887 16248951 16289303 16295790 16295791 16295792 16295793 16295712 16289181 16289182 16288272 16288273 16236900 16297967 16234115 16230792 16261546 16287834 16290673 16290674 16290675 16290676 16290677 16258456 16283904 16283905 16283906 16283908 16283909 16283910 16283911 16283912 16283913 16283889 16273795 16273796 16273797 16273798 16273799 16273800 16283914 16283915 16273811 16273812 16273813 16283916 16283890 16283917 16283918 16283919 16283920 16232587 16232589 16232597 16232601 16232602 16295819 16286845 16295525 16295515 16295517 16295518 16295519 16295520 16295521 16295522 16296145 16287903 16287904 16287905 16296128 16293677 16269258 16269259 16250387 16298003 16276538 16276539 16288276 16288277 16248955 16285023 16285024 16285025 16286559 16286893 16288278 16288279 16287012 16295727 16295728 16289258 16295729 16295730 16295731 16289259 16298004 16275553 16286818 16267701 16252098 16248956 16288261 16287940 16287941 16264670 16285037 16285038 16285040 16295733 16295734 16295735 16289291 16289292 16289293 16289294 16289295 16289296 16289297 16289298 16289299 16289300 16289301 16289302 16290720 16290721 16270006 16270018 16270020 16270025 16270037 16270039 16270040 16270041 16283833 16283834 16266972 16266973 16266974 16266975 16266976 16266977 16266978 16266979 16266991 16266992 16266993 16266994 16266995 16266996 16266997 16266998 16266999 16267000 16267001 16267702 16267703 16267704 16267705 16267706 16267707 16267708 16267709 16267710 16267711 16285026 16248959 16263398 16289185 16289186 16289187 16289188 16289189 16283934 16283936 16283940 16283826 16273429 16266101 16281652 16281653 16285801 16289183 16289184 16290618 16290621 16299622 16299626 16299627 16296223 16296875 16296876 16296877 16289211 16289212 16293848 16287933 16289215 16289220 16287934 16287935 16296878 16289227 16293849 16289228 16289229 16293850 16296879 16293851 16293852 16289231 16289233 16296880 16296881 16296882 16296883 16270967 16276540 16287937 16293678 16290693 16277445 16277503 16293853 16293854 16293860 16293861 16293862 16283835 16283836 16283837 16283838 16283839 16296129 16296130 16285027 16288348 16287872 16293681 16270876
|
| FEI Number |
2016493
|
Recalling Firm/ Manufacturer |
CareFusion 303, Inc. 10020 Pacific Mesa Blvd San Diego CA 92121-4386
|
Manufacturer Reason for Recall | Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications. |
FDA Determined Cause 2 | Device Design |
| Action | On February 14, 2023 and January 9. 2024 Carefusion issued a Urgent medical Device Product Advisory" Recall Notification to affected consignees. Carefusion asked consignees take the following actions:
1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures.
2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them.
3. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter.
4. If you have specific questions about this letter, please contact 1-866-583-8783.
5. Any adverse health consequences experienced with the use of this product should be reported to BD and may be reported to the FDA s MedWatch Adverse Event Reporting
Actions to be taken by distributor
1. Please provide a copy of this notification to any customers who may have received products listed in the Affected Products table above.
2. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. |
| Distribution | US: Nationwide
OUS:JP IT IN DE GB KR NL TW ES BE AU BH CA BR NZ FR AR QA MX CL TR SA MY CH IL IE TH SG KW MA FI AE EG BS BM GR MC PL LB MO IQ TN NO PH CN HK CO JO OM
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|