 Class 2 Device Recall OptiCross HDsee related information
Date Initiated by FirmOctober 22, 2026
Date PostedNovember 20, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0510-2027
Recall Event ID 100133
510(K)NumberK260816 
Product Classification Catheter, ultrasound, intravascular - Product Code OBJ
ProductOptiCross HD Coronary Imaging Catheter, Material Numbers/UPN: H74939352010, H74939352020, H74939352030, H74939352040, H74939352050, H74939352060
Code Information GTIN: 08714729960744, 08714729960751, 08714729960768, 08714729960737, 00191506035419, 00191506035426; All non-expired batches
FEI Number 3002095335
Recalling Firm/
Manufacturer
Boston Scientific Corporation
1 Scimed Pl
Maple Grove MN 55311-1565
For Additional Information ContactClara Johnson
763-494-1133
Manufacturer Reason
for Recall
IFU updates in response to reports of air embolism events and reports of repeated flushing to restore image quality while the catheter was positioned within the patient. Intravascular air may cause air embolism, resulting in clinical consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury.
FDA Determined
Cause 2
Use error
ActionAn Important Medical Device Advisory dated 10/22/26 was sent to customers. Required Action "Review IFU updates related to air embolism as detailed in Appendix 1. "Follow these steps when using OptiCross IVUS catheters: oComplete all device preparation and flushing instructions before inserting the catheter into the patient. This includes clearing air from all accessories and then the catheter by completing all three IFU-prescribed flushing steps. oDo not flush the OptiCross IVUS catheter while it is positioned within the patient if entrapped air is suspected. If additional flushing is required, remove the catheter from the patient before flushing. "Forward this notice to relevant personnel and any facilities to which the product was transferred. If you are a distributor, forward this notice to your customers. "Post this information in a visible location near the affected products to ensure it is readily available to all users of the device. "Complete and return the enclosed Acknowledgement Form. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.
Quantity in Commerce1,288,662 total units
DistributionWorldwide distribution - US Nationwide and Territories and the countries of EMEA, Canada, Japan, LATAM and Asia Pacific countries/regions.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.