Model Number CD3211-36
Device Problem Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/28/2017
Event Type  malfunction  
Event Description
During a device exchange procedure due to normal eri, it was noted that there was difficulty removing the lead from the can due to the set screw.The physician used silicon oil to remove the set screw and connect to the new icd with good electrical measurements.The patient is in stable condition.
 
Manufacturer Narrative
The vtip and vring set screws were not returned along with the device from the field.Testing with sample set screws showed that the vtip and vring connector blocks were able to secure the test leads in the lab.The v is1 lead bore diameters were measured and were found to be within specifications.Since the vtip and the vring set screws were not returned, the cause of the field event remains undetermined.Longevity calculations showed that battery depletion was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE PLUS CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6594220
MDR Text Key76062332
Report Number2938836-2017-24754
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2014
Device Model NumberCD3211-36
Device Lot Number3885604
Other Device ID Number05414734503693
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-