MAUDE Adverse Event Report: MODERN MEDICAL EQUIP MFG LTD SUCTION COAG, F/S, 10FR, 25/CS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MODERN MEDICAL EQUIP MFG LTD SUCTION COAG, F/S, 10FR, 25/CS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 130187
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 12/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.H3 other text : device not yet received.
 
Event or Problem Description
The distributor reported on behalf of their customer that the 130187, suction coag, f/s, 10fr, 25/cs, was being used during an unknown procedure on (b)(6) 2023 when it was reported, ¿¿.The burn caused by the conmed suction coagulator.In order to see the defect i had to use our lighted magnifying glass from our spd department.In picture i show when the malleable tip was bent and on the under surface is a tiny window where there is no insulation.¿ there was no report of medical intervention or hospitalization for the patient or user.Further assessment questions were sent to the reporter; however, to date there has not been a response.This report is being raised due to the reported injury of unknown degree of burn to patient.
 
Additional Manufacturer Narrative
Reported event of burn to patient is confirmed based on photographic evidence provided.Received one 130187 in opened original package.Lot number was verified.Performed a visual inspection and there was no visual abnormalities to the insulation of the handpiece.Root cause cannot be determined, however, based upon ifu; a possible cause of this event could be using a higher recommended setting than what was needed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised the following: use lowest possible power setting on the associated electrosurgical unit capable of achieving desired surgical effect.Activation time should be as short as possible.Do not directly secure the cable with metal instruments to surgical drapes.Activation when in contact with metal instruments may cause burns at the tissue / instrument interface.To avoid burns never allow cable associated with this device to be in contact with skin of patient or touching operator do not permit the cables connected to these devices to be parallel and in close proximity to the leads of other electrical devices.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
The distributor reported on behalf of their customer that the 130187, suction coag, f/s, 10fr, 25/cs, was being used during an unknown procedure on (b)(6) 2023 when it was reported, "the burn caused by the conmed suction coagulator.In order to see the defect i had to use our lighted magnifying glass from our spd department.In picture i show when the malleable tip was bent and on the under surface is a tiny window where there is no insulation." there was no report of medical intervention or hospitalization for the patient or user.Further assessment questions were sent to the reporter; however, to date there has not been a response.This report is being raised due to the reported injury of unknown degree of burn to patient.Updated reporter information: (b)(6) 2024, surgery, tonsillectomy/adenoidectomy, the procedure was completed.Surgeon use a valleylab force fx for cautery.The surgeon has the cut at 8 and the coag at 10.Burn was observed prior to surgery completion.Antibiotic ointment applied and given to patient to use.The user had bent the tip prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUCTION COAG, F/S, 10FR, 25/CS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MODERN MEDICAL EQUIP MFG LTD
unit a, 10/f mai wah ind bldg
1-7 wah sing street
kwai chung, nt
HK 
Manufacturer (Section G)
MODERN MEDICAL EQUIP MFG LTD
unit a, 10/f mai wah ind bldg
1-7 wah sing street
kwai chung, nt
HK  
Manufacturer Contact
john berga
11311 concept blvd
largo, FL 33773
7273995358
MDR Report Key18491229
Report Number3007305485-2024-00008
Device Sequence Number17060336
Product Code GEI
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
K033003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number130187
Device Lot Number202309044
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2023
Supplement Date Received by Manufacturer02/27/2024
Initial Report FDA Received Date01/11/2024
Supplement Report FDA Received Date02/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VALLEYLAB FORCE FX
Outcome Attributed to Adverse Event Other;
Patient Age4 YR
Patient SexMale
Patient Weight21 KG
Patient EthnicityNon Hispanic
-
-