|
As reported by our canadian affiliates, during implant of a 29'mm sapien 3 ultra resilia in the aortic position via a right transfemoral artery, resistance was encountered during insertion of the delivery system when exiting the tip of the esheath+.Once the delivery system reached the tip of the sheath, a distal tip tear to the esheath+ was noted and a portion of the distal tip was found to be missing.The valve was successfully implanted.The esheath+ was subsequently removed without difficulty.The patient remained in stable condition throughout the procedure.It is believed the tip of the esheath+ could still be in the patient, however there is no plan to follow the patient.An unsuccessful effort was made to locate the tip using fluoroscopy.The sterile field was also searched, and the tip was never located.Provided imagery shows the esheath liner expanded and the distal tip torn.Tip material was observed to be missing.
|