• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2 PLUS; CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2 PLUS; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 78747-01
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Codes Hypoglycemia (1912); Diaphoresis (2452)
Date of Event 03/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A high glucose reading issue was reported with the abbott diabetes care (adc) device.The customer received an unspecified high glucose reading on the adc device compared to an unspecified glucose reading obtained on the built-in precision meter.The customer noted a horizontal trend arrow which indicates glucose is changing slowly (less than 1 mg/dl per minute).As a result, the customer experienced cold sweats and was unable to self-treat, requiring third-party administration of juice and dark chocolate for treatment.There was no report of death or permanent impairment associated with this event.A higher reading of 8.6 mmol/l on the adc device was compared to the built-in precision meter reading of 4.8 mmol/l on the same day of sensor wear.The results, when plotted on a parkes error grid, fall into the c zone, showing the difference in values to be clinically significant.It is unknown when these readings were obtained in relation to the reported adverse event.
 
Additional Manufacturer Narrative
The product has been requested back for investigation and the customer agreed to return the device.A follow-up report will be submitted upon completion of investigation activities or if additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A high glucose reading issue was reported with the abbott diabetes care (adc) device.The customer received an unspecified high glucose reading on the adc device compared to an unspecified glucose reading obtained on the built-in precision meter.The customer noted a horizontal trend arrow which indicates glucose is changing slowly (less than 1 mg/dl per minute).As a result, the customer experienced cold sweats and was unable to self-treat, requiring third-party administration of juice and dark chocolate for treatment.There was no report of death or permanent impairment associated with this event.A higher reading of 8.6 mmol/l on the adc device was compared to the built-in precision meter reading of 4.8 mmol/l on the same day of sensor wear.The results, when plotted on a parkes error grid, fall into the c zone, showing the difference in values to be clinically significant.It is unknown when these readings were obtained in relation to the reported adverse event.
 
Additional Manufacturer Narrative
The product has been requested back for an investigation and the customer agreed to return the device; however, at this time product has not yet been received.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the libre sensor and sensor kit was reviewed and the dhrs showed the libre sensor and sensor kit met specifications prior to release to distribution.If product is returned, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.Section d4 (primary udi number) has been updated.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2 PLUS
Common Device Name
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
patricia reis castro
1360 south loop road
alameda, CA 94502-7001
4157205405
MDR Report Key24743499
Report Number2954323-2026-19672
Device Sequence Number17266043
Product Code QBJ
UDI-Device Identifier05021791006991
UDI-Public(01)05021791006991(17)270331(10)251108P
Combination Product (Y/N)N
Initial Reporter CountryHR
PMA/510(K) Number
K223435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number78747-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2026
Supplement Date Received by Manufacturer04/10/2026
Initial Report FDA Received Date03/31/2026
Supplement Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-