Beckman Coulter, Inc ACCESS 2 IMMUNOASSAY ANALYZER; analyzer, chemistry (photometric, discrete), for clinical use
|
Back to Search Results |
|
| Catalog Number |
81600N |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A1: the full patient identifier is (b)(6).A2, a3, a4, a5 and a6: the customer did not supply patient demographics such as age, date of birth, sex, weight, ethnicity or race.H3 and h6: as part of the investigation, the lss corrected the hstni sample unit configuration.After the sample units were modified to the appropriate setting, hstni results returned to expected values, confirming normal assay performance.Beckman coulter product security was also consulted regarding the modified apf software; however, the review did not confirm a product cybersecurity event or incident.The investigation concluded that the cause of the event was user error, as the customer did not ensure consistency between plasma and serum sample unit settings and did not recognize the unit discrepancy when reviewing results.The issue was resolved through correction of the assay unit settings, the customer accepted the resolution, and no further action was required.Retrospective mdr submission under capa-001570.The complaint was reassessed using beckman coulter's revised mdr malfunction reportability methodology and was determined to be reportable.This report is being submitted as part of capa-001570 remediation activities.
|
| |
|
Event or Problem Description
|
|
The customer reported receiving unexpectedly high access hstni (part number b52699) patient results on an access 2 analyzer (serial number (b)(6)).No adverse patient impact, injury, exposure, misdiagnosis, or mistreatment was reported in connection with this event.Following the complaint, a beckman coulter laboratory solution specialist (lss) was dispatched to investigate the issue.During troubleshooting, the lss learned that the elevated hstni results occurred after testing a bnp assay that utilized a third-party reagent.Further investigation determined that the apf software modified by the third-party manufacturer was being used for bnp testing.The bnp assay was configured to use plasma units (pg/ml), while the hstni assay was configured to use serum units (ng/ml).The investigation found that after bnp testing was performed, the hstni assay unit setting changed from serum units (ng/ml) to plasma units (pg/ml).The customer did not recognize this unit change when reviewing results and did not convert the plasma-based results to the appropriate serum units, leading to the perception of falsely elevated hstni results.No hardware errors, assay-specific performance issues, or other system-related problems were reported.System performance data, including system checks, calibration, and quality control results at the time of the event, were not available for review.
|
| |
|
Search Alerts/Recalls
|
|
|