MAUDE Adverse Event Report: Philips Medical Systems Nederland B.V. IntelliSpace Cardiovascular; Cardiovascular Image and Information Management System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Philips Medical Systems Nederland B.V. IntelliSpace Cardiovascular; Cardiovascular Image and Information Management System Back to Search Results
Model Number IntelliSpace Cardiovascular
Medical Device Problem Code Computer System Security Problem (2899)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
Philips identified a software defect internally in intellispace cardiovascular (iscv) whereby the thick client application does not adequately distinguish between legitimate and malicious dynamic link library (.Dll) files.A malicious dll could potentially be placed in a higher-priority search location and subsequently loaded and executed by the application when a dll with a matching name is found.The device was not in clinical use at the time of the event, and no adverse events or harm were reported in the complaint (pr (b)(4).\ \ the investigation revealed that an exploitation of this weakness could allow the attacker to further compromise iscv thick client, deploy malware, or make iscv client unavailable.If iscv behavior is altered, users may be shown incorrect, incomplete, or misleading information and in worst case also make iscv unavailable.\ \ a risk assessment was performed and determined that this affects core clinical workflow functions that depend on timely access to accurate patient information.The issue is not readily recoverable by clinicians, and no workaround exists to maintain system availability or verify data integrity following exploitation.The resulting loss of access to clinical information or use of incorrect patient data may potentially lead to a delayed diagnosis.Based on the available information, this issue is determined to be a reportable malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IntelliSpace Cardiovascular
Common Device Name
Cardiovascular Image and Information Management System
Manufacturer (Section D)
Philips Medical Systems Nederland B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
Philips Medical Systems Nederland B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
edita reznik-shmueli
veenpluis 6
best 5684 -PC
NL   5684 PC
MDR Report Key51119758
Report Number3000976525-2026-100000
Device Sequence Number9981473
Product Code LLZ
UDI-Device Identifier00884838125773
UDI-Public(01)00884838125773(11)251114(10)9.2.0.0
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K251215
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIntelliSpace Cardiovascular
Device Catalogue Number830089
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2026
Initial Report FDA Received Date10/16/2026
Is the Device Labeled for Single Use? (Y/N) No
Patient Sequence Number1
Patient SexUnknown
-
-