Philips identified a software defect internally in intellispace cardiovascular (iscv) wherein the application uses an outdated version of asp.Net in the service tool, which may expose to known security vulnerabilities.The device was not in clinical use at the time of the event, and no adverse events or harm were reported in the complaint ((b)(6)).\
\
the investigation revealed that this creates a potential vulnerability that could be exploited to gain unauthorized access to the service tool environment.Successful exploitation could allow an attacker to modify service tool configurations, potentially resulting in unintended system behavior and impairment of iscv availability, either across the entire system or within specific functions.\
\
a risk assessment was performed and determined that exploitation this vulnerability could impair the availability or expected operation of iscv, either across the entire system or within specific functions.If such loss of availability occurs during clinical use, clinicians may be unable to access or process safety-critical information in a timely manner, potentially delaying clinical review, decision-making, or intervention.Based on the available information, this issue is determined to be a reportable malfunction.
|