Beckman Coulter, Inc UNICEL DXI 800 ACCESS IMMUNO ANALY; Analyzer, chemistry (photometric, discrete), for clinical use
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a sample presentation unit (spu) motion error on a unicel dxi 800 access immunoassay analyzer, product number 973100, serial number (b)(6).As reported, patient samples could not be tested while the issue was present.No erroneous patient results were generated or reported.The customer indicated that patient result reporting was delayed because no backup instrument was available; however, no delay in patient treatment or diagnosis was reported.\
\
there was no injury, death, or delay/change in treatment or diagnosis related to the reported issue and there was no report of harm for a patient, an operator or any other person.
|
| |
|
Additional Manufacturer Narrative
|
|
A1: the full patient identifier is (b)(4).\
\
a2, a3, a4, a5 and a6: the customer did not supply patient demographics such as age, date of birth, sex, weight, ethnicity or race.\
\
a dealer service engineer was dispatched to the customer site for evaluation and troubleshooting.Investigation confirmed the reported spu motion error.The engineer adjusted the spu position as a routine corrective action.No parts were replaced.The adjustment resolved the issue, and the instrument was returned to normal operation.\
\
during follow-up investigation, it was determined that the instrument was operating with non-beckman coulter substrate, wash buffer, and reagent materials (bnp and pct).It was also reported that the apf software had been modified by orienter biotech.The information was forwarded to beckman coulter product security for assessment.The subsequent review did not confirm the software modification as a product cybersecurity event or incident.\
\
conclusion: the investigation identified evidence of a hardware malfunction associated with the reported spu motion error.Adjustment of the spu position resolved the issue, and the instrument remained in normal use following service.\
\
retrospective mdr submission under capa-001570.The complaint was reassessed using beckman coulter's revised mdr malfunction reportability methodology and was determined to be reportable.This report is being submitted as part of capa-001570 remediation activities.
|
| |
|