Model Number N/A
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported a sample presentation unit (spu) motion error on a unicel dxi 800 access immunoassay analyzer, product number 973100, serial number (b)(6).As reported, patient samples could not be tested while the issue was present.No erroneous patient results were generated or reported.The customer indicated that patient result reporting was delayed because no backup instrument was available; however, no delay in patient treatment or diagnosis was reported.\ \ there was no injury, death, or delay/change in treatment or diagnosis related to the reported issue and there was no report of harm for a patient, an operator or any other person.
 
Additional Manufacturer Narrative
A1: the full patient identifier is (b)(4).\ \ a2, a3, a4, a5 and a6: the customer did not supply patient demographics such as age, date of birth, sex, weight, ethnicity or race.\ \ a dealer service engineer was dispatched to the customer site for evaluation and troubleshooting.Investigation confirmed the reported spu motion error.The engineer adjusted the spu position as a routine corrective action.No parts were replaced.The adjustment resolved the issue, and the instrument was returned to normal operation.\ \ during follow-up investigation, it was determined that the instrument was operating with non-beckman coulter substrate, wash buffer, and reagent materials (bnp and pct).It was also reported that the apf software had been modified by orienter biotech.The information was forwarded to beckman coulter product security for assessment.The subsequent review did not confirm the software modification as a product cybersecurity event or incident.\ \ conclusion: the investigation identified evidence of a hardware malfunction associated with the reported spu motion error.Adjustment of the spu position resolved the issue, and the instrument remained in normal use following service.\ \ retrospective mdr submission under capa-001570.The complaint was reassessed using beckman coulter's revised mdr malfunction reportability methodology and was determined to be reportable.This report is being submitted as part of capa-001570 remediation activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXI 800 ACCESS IMMUNO ANALY
Common Device Name
Analyzer, chemistry (photometric, discrete), for clinical use
Manufacturer (Section D)
Beckman Coulter, Inc
1000 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
Beckman Coulter, Inc
1000 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
shama agtey
1000 lake hazeltine drive
chaska, MN 55318
+1(952)491-6938
MDR Report Key51234828
Report Number2122870-2026-00856
Device Sequence Number9636335
Product Code JJE
UDI-Device Identifier15099590265373
UDI-Public(01)15099590265373(11)190627
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K023764
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number973100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2025
Initial Report FDA Received Date10/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-