The device referenced in this report has not yet been returned to olympus for evaluation.The exact cause of the patient's outcome could not be conclusively determined at this time.However, review of the instrument history showed that the device was purchased on (b)(6) 2013, but the device has not been returned to olympus for service since then.If additional information becomes available at a later time, this report will be supplemental.Please reference the associates devices, see mfr reports: 2951238-2014-400, 401,402, 403, 404, 405, 406, 407, 408, 409, 410, 411, 412, 413, 414, 415, 416, 418, 419, 420, 421.
|
This supplemental report is being submitted to provide additional information regarding the in-service visits provided by the endoscopy support specialist (ess) on august 21, 2014 and october 3, 2014.The ess visited the user facility to observe the user facility's reprocessing practices.During the in-service visit it was observed that the endoscopes were being hung with peel pack covers (distal tip protectors) over the insertion tubes while drying.As part of educating the customer, a copy of the distal tip protector letter from infection control was provided to the user facility as well as both manuals on the proper cleaning and reprocessing of the 180 series bronchoscopes and colonoscopes.It was also reported that no other reprocessing deviations were found during the on-site visit.Olympus does not recommend nor endorse the use of the distal tip protectors during storage.It was noted that the endoscope distal tip protectors are not to be reprocessed or reused.Reuse could present an infection control risk if the distal end protectors cannot be thoroughly dried.Additionally, the attachment of a non-sterile distal tip protector to the endoscope which has been high-level disinfected or sterilized may cause recontamination of the endoscope.Olympus bronchoscope instruction manual advises users: "storage of endoscope: to hang the endoscope in the storage cabinet with the distal end hanging freely." if additional or significant information is available at a later time, this report will be supplemented accordingly.
|