CYBERONICS INC PULSE GEN MODEL 102 GENERATOR Back to Search Results
Model Number 102
Event Date 01/05/2015
Event Type  Injury  
Event Description

It was reported that the vns patient had six seizures in one week in 25 minutes and that the magnet was not aborting the seizures. The patient was subsequently admitted to the hospital. The patient was reported to have a uti and a high intake in soda which may have triggered the seizures. Review of the available programming and diagnostic history showed normal diagnostic results through (b)(6) 2007. A battery life calculation using the available programming history showed approximately 2. 6 years remaining. Attempts for additional relevant information have been unsuccessful to date.

 
Manufacturer Narrative

Review of the available programming and diagnostic history. Device manufacturing records were reviewed. Review of manufacturing records confirmed that the generator passed all functional tests prior to distribution.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePULSE GEN MODEL 102
Type of DeviceGENERATOR
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4485095
Report Number1644487-2015-03755
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial
Report Date 01/12/2015
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received02/04/2015
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator LAY USER/PATIENT
Device EXPIRATION Date08/30/2008
Device MODEL Number102
Device LOT Number016158
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received01/12/2015
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured10/10/2006
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 02/04/2015 Patient Sequence Number: 1
-
-