1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE CONV FB PS TB TRL SZ5; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500975
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Date 01/22/2015
Event Type  malfunction  
Event Description
The articular surface cracked and broken.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: unavailable.Follow-up with the complainant has been conducted for the catalog number and lot number and this information is unavailable.
 
Manufacturer Narrative
Additional narrative: examination of the returned device confirms the reported event of trial breakage.A complaints review was completed in february 2014 and in june 2014 for all attune trial instruments, following 25,000 and 50,000 surgeries, inclusive of the attune femoral trials, articulation surface trials, and the patella trials.Findings of the complaints review prompted dfmea updates where applicable.The root cause is attributed to user technique and/or misuse.The damage suggests the trial was pryed or otherwise inappropriately removed during trialing.Based on the root cause of suspected misuse, corrective action is not needed.Monitor subsequent reports via (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV FB PS TB TRL SZ5
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4508644
MDR Text Key5441946
Report Number1818910-2015-13280
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500975
Device Lot NumberBFA0LLB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/02/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/07/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-