• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE RP CR ARTIC SURF SZ7; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE RP CR ARTIC SURF SZ7; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500567
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem No Information (3190)
Event Date 02/25/2015
Event Type  malfunction  
Event Description
The attune articulate surface broke while trialing.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of trial breakage.A complaints review was completed in february 2014 and in june 2014 for all attune trial instruments, following 25,000 and 50,000 surgeries, inclusive of the attune femoral trials, articulation surface trials, and the patella trials.Findings of the complaints review prompted dfmea updates where applicable.The root cause is attributed to user technique and/or misuse.The damage suggests the trial was pryed or otherwise inappropriately removed during trialing.Based on the root cause of suspected misuse, corrective action is not needed.Monitor subsequent reports via sep-419.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
The investigation remains closed, as the added information does not change the outcome of the investigation.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP CR ARTIC SURF SZ7
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4567919
MDR Text Key5474888
Report Number1818910-2015-15171
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK140881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor,distributor,health prof
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500567
Device Lot NumberMVMBKL170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received08/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age50 YR
Patient Weight88
-
-