• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE FEMORAL IMPACTOR; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE FEMORAL IMPACTOR; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254401006
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 02/28/2015
Event Type  malfunction  
Event Description
Attune femoral impactor cracked while impacting the femoral implant.Update-02/28/2015 product received.Upon receipt of the product, it was noted to be broken and not cracked, changing the mdr decision.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Conclusion and justification status: the complaint states attune femoral impactor cracked while impacting the femoral implant.The investigation confirmed that the impactor had broken.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune femoral impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.The annealed product was released on 30-jul-2014.This device is from an un-annealed batch.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that (b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEMORAL IMPACTOR
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4604040
MDR Text Key5630085
Report Number1818910-2015-16213
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Service and Testing Personnel
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401006
Device Lot NumberAU3704961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received03/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age52 YR
-
-