• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE FB TIBIAL IMPACTOR; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE FB TIBIAL IMPACTOR; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254401003
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Type  malfunction  
Event Description
The attune femoral impactor is cracked.The attune tibial impactor is broken into pieces.The twist knob broke off of the attune femoral introducer.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The complaint states the attune femoral impactor is cracked.The attune tibial impactor is broken into pieces.The twist knob broke off of the attune femoral introducer.The investigation could not confirm the above statement as no products were returned.Broken expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune impactors have been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that (b)(4) has been initiated and can be referenced for further details.The complaint mentions the impactors cracked/broke but there was no delay to surgery.Both failure modes; patient harm & delay to surgery have been adequately covered in the risk assessment of the device and no further updates are required.The dfmea scores reflect the current failure rates of the device (b)(4).The complaint shall be closed with an undetermined conclusion; it shall be entered onto the complaints database and monitored through trend analysis.Should further information be made available, then the complaint shall be reviewed and further investigation completed as necessary.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIBIAL IMPACTOR
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4622158
MDR Text Key5589436
Report Number1818910-2015-16798
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401003
Device Lot NumberAU3653985
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/23/2015
Initial Date FDA Received03/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/27/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-