The complaint states the attune femoral impactor is cracked.The attune tibial impactor is broken into pieces.The twist knob broke off of the attune femoral introducer.The investigation could not confirm the above statement as no products were returned.Broken expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune impactors have been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that (b)(4) has been initiated and can be referenced for further details.The complaint mentions the impactors cracked/broke but there was no delay to surgery.Both failure modes; patient harm & delay to surgery have been adequately covered in the risk assessment of the device and no further updates are required.The dfmea scores reflect the current failure rates of the device (b)(4).The complaint shall be closed with an undetermined conclusion; it shall be entered onto the complaints database and monitored through trend analysis.Should further information be made available, then the complaint shall be reviewed and further investigation completed as necessary.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|