Catalog Number 254401004 |
Device Problem
Break (1069)
|
Patient Problem
No Information (3190)
|
Event Date 02/23/2015 |
Event Type
malfunction
|
Event Description
|
Attune fb impactor is cracked.Update-03-10-2015-product received.Upon examination of the product, it was determined the instrument was broken, changing the mdr decision.The complaint was updated 03-12-2015.
|
|
Manufacturer Narrative
|
The complaint states attune fb impactor is cracked.The investigation confirmed that the impactor had broken.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune impactors have been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that (b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|