• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDICS ¿ REG. ATTUNE FB CR ARTIC SURF SZ6; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDICS ¿ REG. ATTUNE FB CR ARTIC SURF SZ6; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500526
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 05/09/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint is still under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Event Description
Location lug has snapped off.
 
Manufacturer Narrative
This complaint is under investigation.Depuy will notify the fda of the results of the investigation once it is completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The investigation remains closed, as the added information does not change the outcome of the investigation.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of trial breakage.A complaints review was completed in february 2014 and in june 2014 for all attune trial instruments, following 25,000 and 50,000 surgeries, inclusive of the attune femoral trials, articulation surface trials, and the patella trials.Findings of the complaints review prompted dfmea updates where applicable.The root cause is attributed to user technique and/or misuse.The damage suggests the trial was pryed or otherwise inappropriately removed during trialing.Based on the root cause of suspected misuse, corrective action is not needed.Monitor subsequent reports via sep-419.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB CR ARTIC SURF SZ6
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDICS ¿ REG.
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDICS ¿ REG.
325 paramount drive
raynham MA 02767
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4771902
MDR Text Key16341407
Report Number1818910-2015-20726
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK140881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,foreign,health professi
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 05/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500526
Device Lot NumberMVMBSL600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/12/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-