• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE CR FEM TRIAL SZ 7 LT; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE CR FEM TRIAL SZ 7 LT; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500717
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 06/10/2015
Event Type  malfunction  
Event Description
The attune trial broke in two pieces.
 
Manufacturer Narrative
(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
The investigation remains closed, as the added information does not change the outcome of the investigation.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of trial breakage.A complaints review was completed in february 2014 and in june 2014 for all attune trial instruments, following 25,000 and 50,000 surgeries, inclusive of the attune femoral trials, articulation surface trials, and the patella trials.Findings of the complaints review prompted dfmea updates where applicable.Previous evaluations by depuy material science for similar failures on the attune femoral trial family found the fracture of the femoral trial was due to overload.No material defects or inclusions were noted that would have contributed to the fracture or propagation.The root cause is attributed to user technique and/or misuse.The damage suggests the trial was pryed or otherwise inappropriately removed during trialing.Based on the root cause of suspected misuse, corrective action is not needed.Monitor subsequent reports via sep-419.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM TRIAL SZ 7 LT
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4840007
MDR Text Key5945675
Report Number1818910-2015-23371
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK140881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor,distributor,health prof
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 06/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number254500717
Device Lot NumberMVMBGZ040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/03/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age61 YR
-
-