Conclusion and justification status: the complaint states during impaction of the tibial tray the attune fb tibial impactor split into two pieces and surgeon discarded it onto the floor.Rep did not notice any evidence of cracking or wear prior to impactor breaking.Unfortunately the hospital has thrown out the broken impactor so it is not available for return.Patient stable on table post surgery.No delay to procedure.No adverse event.The investigation confirmed that the impactor had broken as reported with the photo provided.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune fb tibial impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that (b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
|