• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE SHIM SZ7 5MM; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE SHIM SZ7 5MM; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500671
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 06/15/2015
Event Type  Injury  
Event Description
The size 7/8 5mm attune shim broke when removing after trialing.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Additional narrative: examination of the returned device confirms the reported event of trial breakage.A complaints review was completed in (b)(4) 2014 and in (b)(4) 2014 for all attune trial instruments, following 25,000 and 50,000 surgeries, inclusive of the attune femoral trials, articulation surface trials, and the patella trials.Findings of the complaints review prompted dfmea updates where applicable.The root cause is attributed to user technique and/or misuse.The damage suggests the trial was pryed or otherwise inappropriately removed during trialing.Based on the root cause of suspected misuse, corrective action is not needed.Monitor subsequent reports via (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIM SZ7 5MM
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4865664
MDR Text Key6061828
Report Number1818910-2015-24251
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500671
Device Lot NumberMVMBLC920
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/09/2015
Initial Date FDA Received06/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
Patient Weight140
-
-