Model Number MMT-754WWS |
Device Problems
Device Operates Differently Than Expected (2913); Moisture or Humidity Problem (2986)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 07/13/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
The customer reported via phone call indicating that the insulin pump had keypad anomaly.Customer's blood glucose was 8.6 mmol/l.The customer stated that here is no response from esc button or only on few attempts.The customer stated that the device was exposed to moisture sweat.The customer insulin pump was replaced.
|
|
Manufacturer Narrative
|
The insulin pump had an intermittent button response on the esc button due to corroded keypad traces noted.The insulin pump had a cracked lcd window, cracked case on the lcd window corners, cracked reservoir tube lip, scratches on reservoir tube window and cracked battery tube threads.
|
|