• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-754WWS
Device Problems Device Operates Differently Than Expected (2913); Moisture or Humidity Problem (2986)
Patient Problem Hyperglycemia (1905)
Event Date 07/13/2015
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
The customer reported via phone call indicating that the insulin pump had keypad anomaly.Customer's blood glucose was 8.6 mmol/l.The customer stated that here is no response from esc button or only on few attempts.The customer stated that the device was exposed to moisture sweat.The customer insulin pump was replaced.
 
Manufacturer Narrative
The insulin pump had an intermittent button response on the esc button due to corroded keypad traces noted.The insulin pump had a cracked lcd window, cracked case on the lcd window corners, cracked reservoir tube lip, scratches on reservoir tube window and cracked battery tube threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
christopher tupper
18000 devonshire street
northridge, CA 91325-1219
8185764313
MDR Report Key4946255
MDR Text Key23015781
Report Number3004209178-2015-78172
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Patient
Report Date 07/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754WWS
Device Catalogue NumberMMT-754WWS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received07/13/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-