MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-554WWP
Device Problems Break (1069); Computer Operating System Problem (2898)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/18/2015
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that the customer had a serial peripheral interface to motor programmable integrated circuit communication error alarm twice this week.Customer's blood glucose was 9.3 mmol/l.Customer also had a problem with a prime loop and keypad anomaly.Customer stated that the esc and act buttons were unresponsive.The customer was advised that the insulin pump will need to be replaced.
 
Manufacturer Narrative
A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing. after testing, it was concluded that the device operated within specifications. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
christopher tupper
18000 devonshire street
northridge, CA 91325-1219
8185764313
MDR Report Key4946287
MDR Text Key23049059
Report Number3004209178-2015-78202
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Report Date 07/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-554WWP
Device Catalogue NumberMMT-554WWP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received07/18/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-