The penumbra cat8 indigo system (cat8) was kinked approximately 4.5, 19.5, 21.0, 23.0, 24.0, 26.0, and 37.0 cm from the distal tip.The complaint has been evaluated.The complaint indicated that the cat8 was kinked during the procedure.The procedure was completed with the same catheter.Evaluation of the returned device confirmed the issue mentioned in the complaint.The cat8 was kinked in the proximal shaft.This type of damage typically occurs due to improper handling during use.If the device is manipulated at an angle while force is being applied, damage such as this may occur.It is likely that some of the reported damage may have occurred after the procedure was completed.These devices are 100% visually inspected for damage during process inspection.
|