• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE FEMORAL IMPACTOR; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE FEMORAL IMPACTOR; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254401006
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 09/09/2015
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Attune impactors are broken and missing pieces.
 
Manufacturer Narrative
Conclusion and justification status: the complaint states attune impactors are broken and missing pieces.The investigation confirmed that the impactors had broken as reported.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune impactors have been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that capa-(b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.Depuy still considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEMORAL IMPACTOR
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key5106068
MDR Text Key26828482
Report Number1818910-2015-31621
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Report Date 09/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401006
Device Lot NumberAU3357853
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/25/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-