Catalog Number 254401006 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Event Description
|
Upon routine investigation of consigned instruments it was discovered there are cracks in each of the two impactors.Upon evaluation of the photos provided the attune fb impactor is broken and has large piece missing on the backside of the component.The attune femoral impactor is cracked and there is evidence of a piece missing.
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
Conclusion and justification status: the complaint states upon rountine investigation of consigned instruments it was discovered there are cracks in each of the two impactors.The investigation confirmed that the impactors had cracked and had pieces missing expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune impactors have been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that (b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
|
|