Model Number 302-20 |
Device Problems
Fluid/Blood Leak (1250); Fracture (1260); Mechanical Problem (1384); Low impedance (2285)
|
Patient Problem
No Information (3190)
|
Event Date 08/12/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
|
|
Event Description
|
Patient underwent generator replacement on (b)(6) 2016 and the lead impedance at the time of implant was 687 ohms.The patient had a follow up visit on (b)(6) 2016 and after interrogation, the impedance shown was below 600 ohms, indicating low impedance (< 600 ohms).No known interventions were taken since the patient is receiving efficacy.The impedance will be monitored during each clinic visit.
|
|
Manufacturer Narrative
|
Relevant tests/laboratory data, including dates; corrected data: (b)(6) 2016 09:19:46 am: diagvinitialprechange ¿ 483 ohms / diagvinitialpostchange ¿ 650 ohms / time of change detection ¿ (b)(6) 2016 16:17 estimated occurrence (implant (b)(6) 2016).Initial mdr inadvertently indicated the implant date to be (b)(6) 2016 instead of (b)(6) 2016.
|
|
Event Description
|
Additional programming data was received.
|
|
Event Description
|
Additional programming data was received.
|
|
Event Description
|
Patient underwent lead revision surgery on (b)(6) 2016.The generator was tested with the resistor pin and the impedance was 3246 ohms.The lead wire was noticeably frayed and thus a new lead was implanted.Impedance on new lead when connected to the existing generator was 1029 ohms.The explanted lead was received on 11/28/2016.Analysis is underway but has not been completed to date.
|
|
Event Description
|
Abraded openings were noted on the outer and the inner silicone tubing of the lead coils resulting in portions of the lead coils being exposed.A break was identified in the positive coil.Scanning electron microscopy images of the positive coil broken ends show what appears to be wear (flat surfaces) on the coils strands resulting in reduction of the diameter of the quadfilar coil strands up to the point of break.Also, scanning electron microscopy images of the lead coils at the suspected torn ends of the first portion of the returned lead show appearance suggesting that the coils were exposed to some type of electro-cautery tool (most likely at explant).One strand shows appearance suggesting that the coils were torn.Since the electrode array portion was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.
|
|
Search Alerts/Recalls
|