MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H180
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bacterial Infection (1735)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Olympus cannot determine if the gastroscope referenced in the article has been returned to olympus for evaluation or not, as there was no specific serial number of gastroscope was provided in the article.The cause of the reported incidents cannot be determined at this time, but improper maintenance of the device cannot be ruled out as a contributory factor to the reported events.Olympus will continue to investigate this report and will update the report if additional information becomes available at a later time.
 
Event or Problem Description
On november 1, 2016 olympus received a clinical article titled, ¿carbapenem-resistant klebsiella pneumoniae (cr-kp) cluster associated with gastroscope exposure among surgical intensive care unit (sicu) patients upmc.The article showed that in january 2016 an increase in cr-kp isolates was observed in the open 8 bed sicu at the facility.A total of six cases were identified in january and all six cases were exposed to the same or gastroscope.As a result, the facility conducted an investigation to determine the exposure source as well as the relatedness of cr-kp strains.The user facility collected rectal swabs on all current in-patients exposed to the sicu in january 2016 to assess for cre carriage.Aside from the six cases, no patients exposed to the gastroscopes were still in patient at the time of the investigation.The gastroscope was removed from service and reprocessing records were reviewed.The facility conducted a microbiologic assessment utilizing two scope culture techniques.A boroscopy was performed to examine the scope lumen for defects.The facility staff moved patients out of the sicu to thoroughly clean space and conduct ultraviolet light disinfection.The cr-kp isolates were sent to microbial genome epidemiology laboratory (migel) for pulsed-field gel electrophoresis (pfge) typing and whole genome sequencing (wgs) to assess relatedness.Approx 43 additional patients were reportedly exposed to sicu in january, and 19 out of the 43 were still in house and available for rectal screening.Only 17 out of 19 patients had undergone rectal swabbing, as two patients were discharged prior to obtain the swab.One of the patient¿s rectal swab tested positive for cr-kp but this patient was never exposed to the gastroscope.As a result of the positive rectal swab, the facility continued the weekly screening until no new positive results, and subsequent testing were negative for cre; therefore, the weekly screening was stopped.The article showed that based on the assessment conducted on the gastroscope the following information were reported: no deficiencies identified with scope reprocessing, no organisms identified using either of the scope culturing methods, boroscopy of the scope lumen revealed deep scratches and luminal debris, after boroscopy the scope lumen was removed and replaced with new lumen before returning to service.The gastroscope was reportedly used on a patient with a known cr-kp at the end of (b)(6) 2015.The article showed that a cr-kp transmission likely occurred through a contaminated gastroscope.Report 5 of 6 reports.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to make a correction on the procode from gck to fds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key6137172
Report Number2951238-2016-00902
Device Sequence Number2111190
Product Code FDS
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
PK100584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source literature,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberGIF-H180
Device Catalogue NumberGIF-H180
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/01/2016
Supplement Date Received by Manufacturer02/13/2019
Initial Report FDA Received Date11/30/2016
Supplement Report FDA Received Date03/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-