Catalog Number 254401017 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Attune impactor handle, cat no.(b)(4) ,lot no.Nw152969 broke during tibia preparation and attune artic surface ring(cat no.-(b)94),lot no.Mvmbsy780) came out while removing insert during attune case at medanta the medicity causing a surgery delay of 5 mins to find broken parts, this has happened the fourth time.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary conclusion and justification status: the complaint states attune impactor handle, cat no.(b)(4), lot no.Nw152969 broke during tibia preparation.The investigation confirmed that the lever had broken as reported.It should be noted that a field safety notice was issued stating that to reduce the possibility of leaving fragments in patients to adhere to the ifu which include inspecting the instruments to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure the complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|