Catalog Number 254401006
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary: the investigation confirmed the reported event.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the surgeon was impacting a size 7 cr femur when upon impaction with bone cement already on the prosthesis and femur the 2544-01-006 femoral impactor shattered also breaking the quick release handle lever and spring.Attune impaction handle 2544-01-017 also needs to be replaced, we retrieved all pieces.No surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEMORAL IMPACTOR
Type of Device
ATTUNE INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic dr.
warsaw, IN 46582
5743725905
MDR Report Key7526093
MDR Text Key108728406
Report Number1818910-2018-59993
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295130222
UDI-Public10603295130222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401006
Device Lot NumberAU5895371
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age66 YR
-
-