• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FEMORAL IMPACTOR; ATTUNE INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FEMORAL IMPACTOR; ATTUNE INSTRUMENTS : IMPACTORS Back to Search Results
Catalog Number 254401006
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/11/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary: the investigation confirmed that the reported event.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that upon inspection the following instrument was cracked and had a portion of it chipped off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEMORAL IMPACTOR
Type of Device
ATTUNE INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic dr.
warsaw, IN 46582
5743725905
MDR Report Key7573535
MDR Text Key110237123
Report Number1818910-2018-61487
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295130222
UDI-Public10603295130222
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401006
Device Lot NumberAU3704961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/11/2018
Initial Date FDA Received06/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-