• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE MEDIAL ANAT PAT 32MM; ATTUNE IMPLANT : KNEE PATELLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE MEDIAL ANAT PAT 32MM; ATTUNE IMPLANT : KNEE PATELLA Back to Search Results
Model Number 1518-10-032
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problems Pain (1994); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
On (b)(6) 2018 medical records received.After review of the medical records for mdr reportability, the patient was revised to address pain and persistent effusions, patellar maltracking and rotational malalignment of the femoral component, loosening of the tibial tray at the tray/cement interface.Unknown cement used.There is no new additional information that would affect the existing investigation.
 
Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
There was no reported product problem concerning the patella.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : product code 151810032, work order (b)(4) was manufactured on 29-jan-13, (b)(4) parts were manufactured as per specification and all raw materials met specification.There were no ncs or deviations associated with this lot.Due to no similar failures found in the dhr review, the root cause of the complaint cannot be determined.Should further information come available that impacts the findings in this investigation it will be reopened.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
H6 patient code: no code available (3191) used to capture the surgical intervention.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Added: a2, d4, d11, h4, h6.Corrected: a1, g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE MEDIAL ANAT PAT 32MM
Type of Device
ATTUNE IMPLANT : KNEE PATELLA
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork IN
EI 
MDR Report Key7600502
MDR Text Key111088099
Report Number1818910-2018-62152
Device Sequence Number1
Product Code JWH
UDI-Device Identifier10603295056621
UDI-Public10603295056621
Combination Product (y/n)N
PMA/PMN Number
K103756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2017
Device Model Number1518-10-032
Device Catalogue Number151810032
Device Lot Number3568868
Was Device Available for Evaluation? No
Date Manufacturer Received02/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ATTUNE CR FB INSRT SZ 4 5MM; ATTUNE CR FEM RT SZ 4 CEM; ATTUNE FB TIB BASE SZ 4 CEM
Patient Outcome(s) Required Intervention;
Patient Age60 YR
-
-