MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20E DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20E DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20E
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 06/14/2018
Event Type  malfunction  
Manufacturer Narrative
Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event Description
The customer contacted physio-control to report that the service indicator is present and an event code is logged in the memory of their device.The event code logged in its memory is indicative of a device failure that could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.Additionally, the defibrillator output energy may be reduced by up to approximately 20% from the selected energy level.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
 
Manufacturer Narrative
Physio-control has made multiple attempted to contact the customer to obtain the status of their device.No response has been received from the customer.This device was not returned to physio-control for evaluation.The cause of the reported issue could not be determined.Device not evaluated by manufacturer.
 
Event Description
The customer contacted physio-control to report that the service indicator is present and an event code is logged in the memory of their device.The event code logged in its memory is indicative of a device failure that could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.Additionally, the defibrillator output energy may be reduced by up to approximately 20% from the selected energy level.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20E DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7679906
MDR Text Key113754960
Report Number0003015876-2018-01086
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20E
Device Catalogue Number99507-000002
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/25/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-