• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE MEDIAL DOME PAT 35MM; ATTUNE IMPLANT : KNEE PATELLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE MEDIAL DOME PAT 35MM; ATTUNE IMPLANT : KNEE PATELLA Back to Search Results
Model Number 1518-20-035
Device Problems Use of Device Problem (1670); Malposition of Device (2616); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Synovitis (2094); Joint Disorder (2373); Fibrosis (3167); No Code Available (3191)
Event Date 10/24/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Is a non-healthcare professional.(b)(4).
 
Event Description
Revision due to pain and lack of motion.Doi: (b)(6) 2015; dor: oct 24, 2017; left knee.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary:no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).No device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Per internal procedures, the event information and any investigational inputs received as part of required follow up were reviewed.For this investigation, no immediate action was required as no alleged deficiency with the device was identified.Monitor ¿ exempt per wi-7915 - known possible complication of joint replacement surgery.No device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.The reported event is considered one of the possible complications of joint replacement.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample associated with this report, it was not possible to determine if the device failed to meet specification at the time it was released for distribution.The device associated with this event was used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No product contribution to the reported event was identified.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. h10 additional narrative:   corrected: d4.
 
Manufacturer Narrative
Product complaint #
=
> (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Event Description
Update 17 jun 2019.After review of the medical records for mdr reportability: on (b)(6) 2015, the patient underwent a left knee arthroplasty due to osteoarthritis.Two depuy products were used during the primary operation.The surgeon reported no intra-operative complications.On (b)(6) 2017, the patient underwent a left knee revision due to pain, fibrosis, joint contracture, decreased range of motion, synovitis, and patella baja.No evidence of infection.All components were well-fixed.Depuy products were implanted with competitor cement.On (b)(6) 2018, the patient underwent a left knee revision due to loosening of the tibial tray at the cement to implant interface, bone injury/loss to the femur and tibia, scarring, and adhesions.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy as not synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE MEDIAL DOME PAT 35MM
Type of Device
ATTUNE IMPLANT : KNEE PATELLA
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key8277463
MDR Text Key134085408
Report Number1818910-2019-82285
Device Sequence Number1
Product Code JWH
UDI-Device Identifier10603295056683
UDI-Public10603295056683
Combination Product (y/n)N
PMA/PMN Number
K103756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Model Number1518-20-035
Device Catalogue Number151820035
Device Lot Number8107537
Was Device Available for Evaluation? No
Date Manufacturer Received02/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
-
-