• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 5 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 5 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150600005
Device Problems Loss of or Failure to Bond (1068); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Swelling (2091); Synovitis (2094); Joint Disorder (2373); No Code Available (3191)
Event Date 02/26/2019
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.(b)(4).
 
Event Description
Patient communication received.As verbalized by the patient, "i had an attune knee tibial component put in my knee and it has failed.I had a revision." patient c/o pain and swelling prior to revision.Doi: 2014; dor: (b)(6) 2019; (unk knee).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device was returned.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Investigation methods: was patient affected: yes.Device history reviewed: no.Lot trace obtained: no.Complaints database searched: no.Product checked: no.Label checked: no.Product pulled from stock for inspection: no a review of complaint databases or manufacturing records was not possible as no product details were received.Without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.The device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.The complaint shall be closed with an undetermined conclusion; entered into the complaints database and monitored through trend analysis.Post market surveillance is per (b)(4).If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint # (b)(4).Investigation summary: no device was returned.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: product code 150600005, work order 3616669 was manufactured on 15/may/2013.(b)(4) parts were manufactured per specification and all raw materials met specification.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d10 concomitant med products if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 5 CEM
Type of Device
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork IN
EI 
MDR Report Key8465101
MDR Text Key140337534
Report Number1818910-2019-89021
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
PMA/PMN Number
K101433
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number150600005
Device Lot Number3616669
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/08/2019
Initial Date FDA Received03/29/2019
Supplement Dates Manufacturer Received04/25/2019
05/22/2019
01/17/2020
02/15/2021
Supplement Dates FDA Received04/25/2019
06/13/2019
01/23/2020
02/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ATTUNE CR FB INSRT SZ 6 5MM; ATTUNE CR FEM LT SZ 6 CEM; ATTUNE MEDIAL DOME PAT 35MM; SMARTSET GHV GENTAMICIN 40G; SMARTSET GHV GENTAMICIN 40G
Patient Outcome(s) Required Intervention;
Patient Age73 YR
-
-