Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary = > the device associated with the reported event was not returned.The investigation could not draw any conclusions about the reported event breakage without the device to examine.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot null device history batch null.Device history review null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|