• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 ATTUNE RP TIB BASE SZ 4 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 ATTUNE RP TIB BASE SZ 4 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150610004
Device Problem Loss of or Failure to Bond (1068)
Patient Problems Pain (1994); Scar Tissue (2060); No Code Available (3191)
Event Date 07/31/2018
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The patient underwent a right knee revision due to pain and tibial loosening at the cement to implant interface.The surgeon noted the excision of scar tissue.The tibial tray was easily removed with no cement fixation to the implant.Competitor cement was used during the primary operation.Doi: (b)(6) 2017; dor: (b)(6) 2018, right knee.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).No device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIB BASE SZ 4 CEM
Type of Device
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key8897119
MDR Text Key154429822
Report Number1818910-2019-101191
Device Sequence Number1
Product Code NJL
UDI-Device Identifier10603295042587
UDI-Public10603295042587
Combination Product (y/n)N
PMA/PMN Number
P830055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 07/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number150610004
Device Lot Number8602592
Was Device Available for Evaluation? No
Date Manufacturer Received09/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ATTUNE PS FEM RT SZ 4 CEM; ATTUNE PS RP INSRT SZ4 5MM; COMPETITOR BONE CEMENT; ATTUNE PS FEM RT SZ 4 CEM; ATTUNE PS RP INSRT SZ4 5MM; COMPETITOR BONE CEMENT
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-