• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceCOBAS AmpliPrep/COBAS TaqMan HCV Test v2.0
Generic NameAssay, hybridization and/or nucleic acid amplification for detection of hepatitis c rna,hepatitis c virus
Regulation Number866.3170
ApplicantRoche Molecular Systems, Inc.
4300 Hacienda Dr.
Pleasanton, CA 94588
PMA NumberP060030
Supplement NumberS086
Date Received06/14/2021
Decision Date08/11/2021
Reclassified Date12/22/2021
Product Code MZP 
Advisory Committee Microbiology
Supplement Type30-Day Notice
Supplement Reason Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
Change the manufacturer/supplier of a critical sub-assembly component.
-
-