| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | NeuroPace RNS System |
| Generic Name | Implanted brain stimulator for epilepsy |
| Applicant | NeuroPace, Inc. 455 N Bernardo Ave. Mountain View, CA 94043 |
| PMA Number | P100026 |
| Supplement Number | S056 |
| Date Received | 07/03/2018 |
| Decision Date | 08/01/2018 |
| Product Code |
PFN |
| Advisory Committee |
Neurology |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Modify the PXI Vader automated test equipment (ATE) power-up software to improve yield of the components used to manufacture the RNS® Neurostimulator (model RNS-320) during the custom packaged integrated circuitry (Cassandra) and printed circuit assembly (PCA) testing. Additionally, the Dog Tag test limits for the integrated circuit are being modified in the PXI Vader ATE, Telemetry Only System (TOS) ATE and the PXI Vader Temperature ATE databases to allow flexibility to also test future integrated circuits for product development purposes. |
|
|