| Device | Optune System |
| Generic Name | Stimulator, low electric field, tumor treatment |
| Applicant | Novocure, GmbH Neuhofstrasse 21 Baar 6340 |
| PMA Number | P100034 |
| Supplement Number | S019 |
| Date Received | 09/20/2018 |
| Decision Date | 12/19/2018 |
| Product Code |
NZK |
| Advisory Committee |
Neurology |
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for changes to Optune System carrying case and INE Transducer Array package sizes; Optune System software modifications increasing the secondary alarm threshold for INE Transducer Array partial attachment, correcting data transfer processing following switch from battery to mains power, and eliminating a redundant INE Transducer Array overheating alarm; and NovoTerminal software modifications improving Optune System data transfer processing. |
|
|