| Device | THERASCREEN KRAS RGQ PCR KIT |
| Generic Name | Somatic gene mutation detection system |
| Applicant | QIAGEN GmbH Qiagen Strasse 1 Hilden 40724 |
| PMA Number | P110027 |
| Supplement Number | S003 |
| Date Received | 12/22/2014 |
| Decision Date | 01/16/2015 |
| Product Code |
OWD |
| Advisory Committee |
Pathology |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement CHANGE TO THE SUPPLIERS PURITY TESTING METHOD FOR THE INTERNAL CONTROL TEMPLATE IN YOUR THERASCREEN KRAS RGQ PCR KIT. THE NEW PURITY TESTING WILL USE REVERSED PHASE-ULTRA PERFORMANCE LIQUID CHROMATOGRAPHY (RP-UPLC) WITH A CHANGE IN ACCEPTANCE CRITERION FROM >=90% TO >=80%. |