• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceOvation iX Abdominal Stent Graft System
Generic NameSYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
ApplicantEndologix, LLC
3910 Brickway Blvd.
Santa Rosa, CA 95403
PMA NumberP120006
Supplement NumberS036
Date Received06/09/2020
Decision Date07/09/2020
Product Code MIH 
Advisory Committee Cardiovascular
Supplement TypeSpecial (Immediate Track)
Supplement Reason Labeling Change - Indications/instructions/shelf life/tradename
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
Approval for updating the IFU with the following language:Polymer leaks are a unique potential risk of the Ovation device platform that have been reported post-market. The complications of polymer leakage in to the vasculature have ranged from transient hypotension to severe life-threatening anaphylactoid reactions, tissue necrosis and death. When polymer leaks occur, underfilling of the Ovation iX sealing rings have led to intraoperative Type Ia endoleaks that have required additional therapy. The risk of polymer leak should be carefully considered along with the risks associated with alternative treatment options when making personalized treatment decisions for those individuals who fall within the indicated patient population as defined by the Instructions for Use. This updated language was agreed upon as part of the mitigation strategy of the Class I Recall of the Ovation iX.
-
-