| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | Deep Brain Stimulation (DBS) |
| Generic Name | Stimulator, electrical, implanted, for parkinsonian tremor |
| Applicant | ABBOTT MEDICAL 6901 Prestion Rd. Plano, TX 75024 |
| PMA Number | P140009 |
| Supplement Number | S102 |
| Date Received | 07/08/2024 |
| Decision Date | 08/08/2024 |
| Product Codes |
MHY NHL PJS |
| Advisory Committee |
Neurology |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement supplier-related changes, including 1) an additional raw material supplier of Grade 2 Titanium; and 2) changing the primary dimensional inspection method from the optical gaging product measurement method to the laser inspection method, for the feedthru component within the Eterna Spinal Cord Stimulation Implantable Pulse Generator and Liberta RC Deep Brain Stimulation Implantable Pulse Generator |
|
|