| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM |
| Generic Name | Pump, infusion, insulin, to be used with invasive glucose sensor |
| Applicant | Tandem Diabetes Care, Inc. 11045 Roselle St. San Diego, CA 92121 |
| PMA Number | P140015 |
| Supplement Number | S020 |
| Date Received | 03/01/2017 |
| Decision Date | 08/25/2017 |
| Product Codes |
OYC PQF |
| Docket Number | 17M-5262 |
| Notice Date | 09/05/2017 |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | Panel Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for the use of the t:slim X2 Insulin Pump with the Dexcom G5 Mobile CGM and for modifying the indications for use to include pediatric patients ages 6-11 years and replace adjunctive with non-adjunctive CGM use (i.e., replace fingerstick blood glucose testing for diabetes treatment decisions). This device is indicated as follows:The t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM (“t:slim X2 System”) consists of the t:slim X2 Insulin Pump paired with the Dexcom G5 Mobile Sensor and Transmitter. The t:slim X2 Insulin Pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The t:slim X2 Insulin Pump can be used solely for continuous insulin delivery and as part of the t:slim X2 System to receive and display continuous glucose measurements from the Dexcom G5 Mobile Sensor and Transmitter. The t:slim X2 System also includes continuous glucose monitoring (CGM) indicated for the management of diabetes. The Dexcom G5 Mobile CGM is designed to replace fingerstick blood glucose testing for diabetes treatment decisions. The t:slim X2 System aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments, which may minimize these excursions. Interpretation of the t:slim X2 System results should be based on the trends and patterns seen with several sequential readings over time. The t:slim X2 System is indicated for use in individuals 6 years of age and greater. The t:slim X2 System is intended for single patient use and requires a prescription. The device is indicated for use with NovoLog or Humalog U-100 insulin. |
| Approval Order | Approval Order |
| Summary | Summary of Safety and Effectiveness |
| Labeling | Labeling
|
|
|