• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceFreeStyle Libre Pro Flash Glucose Monitoring System
Generic NameSensor, glucose, invasive
ApplicantAbbott Diabetes Care, Inc.
1360 S. Loop Rd.
Alameda, CA 94502
PMA NumberP150021
Supplement NumberS028
Date Received05/01/2018
Decision Date06/01/2018
Product Code MDS 
Advisory Committee Clinical Chemistry
Supplement Type30-Day Notice
Supplement Reason Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
Updating the foil lid sealing temperature of the assembly process for the Sensor Pack of the FreeStyle Libre Pro Flash Glucose Monitoring System and Freestyle Libre Flash Glucose Monitoring System.
-
-