| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | Axonics Sacral Neuromodulation System |
| Generic Name | Stimulator, electrical, implantable, for incontinence |
| Regulation Number | 876.5270 |
| Applicant | Boston Scientific Corporation 100 Boston Scientfic Way Marlborough, MA 01752 |
| PMA Number | P180046 |
| Supplement Number | S017 |
| Date Received | 06/12/2020 |
| Decision Date | 07/10/2020 |
| Product Code |
EZW |
| Advisory Committee |
Gastroenterology/Urology |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Axonics Charging Device uses a printed circuit board assembly (PCBA) that is currently tested manually using test procedure, 120-0113 (VOL# 27, file #009 of original submission). To support an increase in production, Axonics developed a test fixture (# 200-0198-001, built with software 150-0113-001) that automates the tests that are in the original manual test procedure. A complete traceability of test steps in the original manual test procedure to the proposed automated test fixture requirements is provided |
|
|